BK Medical E14C4t User Manual Download Page 15

E14C4t User Guide

 (16-01265-EN-07)

Puncture and Biopsy Facilities

 

15

 

Figure 11. Biopsy channel bracket and needle guide mounted on E14C4t. 

For Transperineal Puncture

The metal puncture attachment UA1324, shown in Figure 12, is designed for 

transperineal puncture and biopsy. When UA1324 is being used, the dummy channel 

bracket UA1325-w (shown in Figure 3) must be in place.
UA1324 consists of a needle guide and a mounting ring with clamp. The needle guide 

comprises 9 parallel guide channels, spaced 5 mm apart, each with an internal 

diameter of 2.1 mm, suitable for a 14-gauge needle. The guideis parallel to the 

centerline of the transducer.

Note: 

The needle guide can be adjusted 70 mm lengthwise with respect to the 

mounting ring, using the adjustment screw.

Figure 12. Puncture attachment UA1324.

Mounting the Transperineal Puncture Guide

To mount the transperineal puncture attachment, ensure that the dummy channel 

bracket UA1325-w is in place. Pull a sterile transducer cover over the transducer. 

Loosen the clamp on UA1324, and slide the attachment over the tip of the transducer 

until it meets the steel stud on the side of the transducer. The puncture attachment 

should be correctly positioned so that the groove slides easily over the stud. No force 

should be used when attaching the puncture attachment to the transducer.

WARNING

P-w9

Ensure that the channel bracket and needle guide are correctly positioned. Never insert the 

needle guide while the transducer is inside the patient.

Summary of Contents for E14C4t

Page 1: ...For Professional Users Only 16 01265 EN 07 June 2019 E14C4t Transducer User Guide...

Page 2: ...a BK Medical transducer contains information about the date of manufacture BK Medical Customer Satisfaction Input from our customers helps us improve our products and services Your opinions are impor...

Page 3: ...Buttons 9 Changing Orientation 9 Imaging with E14C4t 9 Simultaneous Biplane Imaging 9 Transverse or Sagittal Imaging 10 Endfire Imaging 10 Imaging Without Puncture or Biopsy 10 Adjusting Image Area a...

Page 4: ...4...

Page 5: ...ing and endfire imaging in a single transducer E14C4t is designed for transrectal and transvaginal procedures It is also suitable for elastography and contrast imaging Patient Population The patient p...

Page 6: ...Information Product specifications acoustic output data and data about EMC electromagnetic compatibility for this transducer can be found in the Product Data Sheet and the Technical Data BZ2100 that...

Page 7: ...This can damage the cable and lead to degradation and or corruption of the ultrasound image WARNING AO w1 To avoid tissue damage always keep the exposure level the acoustic output level and the expos...

Page 8: ...e CE marked WARNING T w5 To prevent electrical shock and damage to the transducer the connector pins in the transducer plug must always be completely dry before you connect to a system WARNING GS w4a...

Page 9: ...ton enables the sagittal or endfire array A long press on this button also activates endfire imaging A second button enables the transverse array Changing Orientation To change the orientation of the...

Page 10: ...t enable split screen imaging as there is no defined isocenter in this context When you press the system s Split key tap DUAL two images appear on the screen The endfire image appears and either a tra...

Page 11: ...in both the transverse and sagittal planes Endfire imaging in the sagittal plane Dual combines both the simultaneous biplane and endfire imaging in one biopsy guide All the biopsy guides for transrect...

Page 12: ...A1326 are used for simultaneous biplane imaging Figure 5 Reusable biopsy guide UA1326 The puncture line for UA1322 S UA1322 S14 and UA1326 on E14C4t is shown in Figure 6 WARNING Sterile w1 Single use...

Page 13: ...guide UA1327 The puncture line for UA1323 S and UA1327 on E14C4t is shown in Figure 8 The needle guide is parallel to the centerline of the transducer Figure 8 Illustration of the puncture line for b...

Page 14: ...ile imaging gel over the transducer 3 Slide the needle guide into the biopsy channel bracket 4 Insert the assembled needle guide and biopsy channel bracket into the open channel on the transducer A sm...

Page 15: ...he needle guide can be adjusted 70 mm lengthwise with respect to the mounting ring using the adjustment screw Figure 12 Puncture attachment UA1324 Mounting the Transperineal Puncture Guide To mount th...

Page 16: ...User Guide 16 01265 EN 07 Figure 13 Puncture attachment UA1324 mounted on E14C4t The puncture lines for UA1324 on E14C4t are shown in Figure 14 Figure 14 Illustration of the puncture line for punctur...

Page 17: ...instructions WARNING P w1 Before you startimaging verifythatthetypenumberornameofthetransducerandthetype numberordescriptionof the needleguideyouareusing matchthenumberdisplayedonthe monitor Also make...

Page 18: ...must be followed Within the EU when you discard the transducer you must send it to appropriate facilities for recovery and recycling WARNING Reproc w3 Immediately after use you mustpre cleanthe devic...

Page 19: ......

Page 20: ......

Reviews: