Puncture and Biopsy Facilities

 

11

 

With the Expanded Sector feature you can increase the transverse sector angle from 

a default width of 140

 to 180

.

Figure 4. Dotted lines indicate expanded transverse sector on E14C4t.

Puncture and Biopsy Facilities

Puncture and biopsy are possible with E14C4t. The appropriate puncture attachments 

are illustrated in the following pages with a brief description of their uses and 

operating instructions. The transducer has an open channel into which the 

appropriate puncture accessories fit when the dummy channel bracket (UA1325-

w) has been removed (see Figure 3).

For Transrectal Puncture

With E14C4t, you can perform transrectal puncture and biopsy by imaging in one of 

three ways:

Simultaneous biplane

 (imaging in both the transverse and sagittal planes).

Endfire

 (imaging in the sagittal plane).

Dual

 (combines both the simultaneous biplane and endfire imaging in one 

biopsy guide).

All the biopsy guides for transrectal puncture have a bore diameter of 1.6 mm, 

suitable for 17- and 18-gauge needles.
For transrectal puncture, biopsy guides are available in both non-sterile reusable 

versions and sterile-packed single-use versions. 

Non-Sterile Biopsy Guides (Light Green)

The dummy channel bracket UA1325-w and the reusable biopsy guides (UA1326, 

UA1327 and UA1328) are non-sterile when supplied. This is also the case with 

Puncture attachment UA1324. They must all be reprocessed following the procedure 

in 

Care and Cleaning.

The dummy channel bracket and reusable biopsy guides may be damaged during use 

or processing, so they must be checked before use for cracks or irregularities in the 

surface. They should also be checked thoroughly once a month following the 

procedure in 

Care and Cleaning.

Summary of Contents for E14C4t

Page 1: ...For Professional Users Only 16 01265 EN 07 June 2019 E14C4t Transducer User Guide...

Page 2: ...a BK Medical transducer contains information about the date of manufacture BK Medical Customer Satisfaction Input from our customers helps us improve our products and services Your opinions are impor...

Page 3: ...Buttons 9 Changing Orientation 9 Imaging with E14C4t 9 Simultaneous Biplane Imaging 9 Transverse or Sagittal Imaging 10 Endfire Imaging 10 Imaging Without Puncture or Biopsy 10 Adjusting Image Area a...

Page 4: ...4...

Page 5: ...ing and endfire imaging in a single transducer E14C4t is designed for transrectal and transvaginal procedures It is also suitable for elastography and contrast imaging Patient Population The patient p...

Page 6: ...Information Product specifications acoustic output data and data about EMC electromagnetic compatibility for this transducer can be found in the Product Data Sheet and the Technical Data BZ2100 that...

Page 7: ...This can damage the cable and lead to degradation and or corruption of the ultrasound image WARNING AO w1 To avoid tissue damage always keep the exposure level the acoustic output level and the expos...

Page 8: ...e CE marked WARNING T w5 To prevent electrical shock and damage to the transducer the connector pins in the transducer plug must always be completely dry before you connect to a system WARNING GS w4a...

Page 9: ...ton enables the sagittal or endfire array A long press on this button also activates endfire imaging A second button enables the transverse array Changing Orientation To change the orientation of the...

Page 10: ...t enable split screen imaging as there is no defined isocenter in this context When you press the system s Split key tap DUAL two images appear on the screen The endfire image appears and either a tra...

Page 11: ...in both the transverse and sagittal planes Endfire imaging in the sagittal plane Dual combines both the simultaneous biplane and endfire imaging in one biopsy guide All the biopsy guides for transrect...

Page 12: ...A1326 are used for simultaneous biplane imaging Figure 5 Reusable biopsy guide UA1326 The puncture line for UA1322 S UA1322 S14 and UA1326 on E14C4t is shown in Figure 6 WARNING Sterile w1 Single use...

Page 13: ...guide UA1327 The puncture line for UA1323 S and UA1327 on E14C4t is shown in Figure 8 The needle guide is parallel to the centerline of the transducer Figure 8 Illustration of the puncture line for b...

Page 14: ...ile imaging gel over the transducer 3 Slide the needle guide into the biopsy channel bracket 4 Insert the assembled needle guide and biopsy channel bracket into the open channel on the transducer A sm...

Page 15: ...he needle guide can be adjusted 70 mm lengthwise with respect to the mounting ring using the adjustment screw Figure 12 Puncture attachment UA1324 Mounting the Transperineal Puncture Guide To mount th...

Page 16: ...User Guide 16 01265 EN 07 Figure 13 Puncture attachment UA1324 mounted on E14C4t The puncture lines for UA1324 on E14C4t are shown in Figure 14 Figure 14 Illustration of the puncture line for punctur...

Page 17: ...instructions WARNING P w1 Before you startimaging verifythatthetypenumberornameofthetransducerandthetype numberordescriptionof the needleguideyouareusing matchthenumberdisplayedonthe monitor Also make...

Page 18: ...must be followed Within the EU when you discard the transducer you must send it to appropriate facilities for recovery and recycling WARNING Reproc w3 Immediately after use you mustpre cleanthe devic...

Page 19: ......

Page 20: ......

Reviews: