Respiro
43 (48)
Operating Instructions
© Bittium Biosignals 2022
7.5
Electromagnetic emissions
Table 9 Electromagnetic emissions
Manufacturer’s declaration
- Electromagnetic emissions
Respiro is suitable for use in an electromagnetic environment as described below. The users should ensure
that the device is used in such an environment.
Emission test
Compliance
Electromagnetic environment
RF emissions
CISPR11
Group 1
Respiro uses RF energy exclusively for its internal function.
Thus the RF emission is very low and it is unlikely that
nearby electronic devices would be disturbed.
RF emissions
CISPR11
Class B
7.6
Immunity test levels
Table 10 Immunity test levels
Phenomenon
Basic EMC standard or
test method
Immunity test level
Home healthcare environment
Electrostatic discharge
IEC 61000-4-2
±8 kV contact
±2 kV, ±4 kV, ±8 kV, ±15 kV air
Radiated RF EM fields
IEC 61000-4-3
10 V/m
80 MHz
–
2,7 GHz
80 % AM, 1 kHz
Proximity fields from RF
wireless communica-
tions equipment
IEC 61000-4-3
See Table on next page.
RATED power frequency
magnetic fields
IEC 61000-4-8
30 A/m
50 Hz or 60 Hz
Conducted disturbances
induced by RF fields
EC 61000-4-6
3 V
0,15 MHz
–
80 MHz
6 V in ISM and amateur radio bands between
0,15 MHz and 80 MHz