4.9.1.1 Switching off NeurOne System
The Main Unit and Amplifiers powered from Main Unit can be switched off with the
Power Button, by pressing it for over
two seconds
, until the Display backlight switches
off. Power off is possible in all other display modes except FW upgrade (see the figure
in chapter 4.9.1).
Turn off Tesla amplifiers and battery pack(s). If video option is also in use, turn off
video camera and Video Trigger unit.
4.10 Manual activation of calibration files and FW version check
The calibration files (*.hbc) have been copied and activated for use automatically
during calibration and by default they are operational and ready for measurement.
In case of missing calibration file the following procedure is required:
1. By default NeurOne calibration file(s) of the amplifier(s) are located in the
Hbc
subfolder in Misc data folder:
C:\ProgramData\Mega Electronics Ltd\NeurOne64\Hbc
Calibration file name syntax is:
TMS <serial number><date>.hbc
MRI <serial number><date>.hbc
EXG <serial number><date>.hbc
2. If the desired calibration file is not found, please, ask for the calibration file of
desired amplifier.
3. Then copy calibration file to
HBC
subfolder located in defined Misc data folder.
4. Start NeurOne System as instructed in chapter 4.9.
5. In NeurOne PC SW: Go to
Settings
|
Core
|
Manager..
.
6. In the NeurOne Manager topology, move mouse pointer on top of the amplifier
icon, which shows
no Active calibration file
for that amplifier.
7. Press mouse right button on amplifier icon, press
Calibrate … on the pop-up
menu.
8.
Headbox calibration
dialog is opened and calibration file of the amplifier is shown
on the calibration file list.
9. Press Set active. Now the selected Amplifier is calibrated and ready for
measurement.
10. Repeat steps 6
– 9 for all Amplifiers of the system.
11. Check by placing mouse pointer on top of each amplifier icon the firmware
version of the amplifier shown on the tooltip. If needed upgrade the firmware
version as instructed in the chapter 5.8.1.1 NeurOne manager.
The amplifiers with activated calibration file are now ready for EEG measurements.