Pulse Oximeter
Operation
47
1.
Starting from the main menu, call:
Configuration
Acoust.
Alarms
Silence Time
2.
Set the alarm tone silence time
between
30
and
120 sec
.
3.
Confirm this setting with
RETURN
.
NOTICE!
During the alarm silence time, the
display shows the alarm silence
symbol:
.
Fig. 47: Alarm tone silence time
CAUTION! New alarms cannot be detected!
If a new alarm condition occurs during the alarm tone mute duration, the new alarm
will be displayed after completion of the mute duration.
7.6.3 Masimo Signal Processing (
Signalprocessing
)
Starting from the main menu, call:
Configuration
Signalprocessing
Following actual signal processing
parameters can be set:
–
Averaging time (
Averaging time
)
–
Sensitivity (
Algorithm mode
)
–
Smart Tone (
on/off
)
–
Artifact filter (
on/off
)
Fig. 48: Masimo signal processing
7.6.3.1 Averaging time (
Averaging time
)
The averaging time indicates the duration, with which from several original measured values
individual displays one VALUE is again computed.
The averaging time can be adjusted between 2-4 (FastSat
TM)
and 16 seconds. 8 seconds is the
default setting.
The longer the length of time, in which measured values are collected, the lower the display value
varies.