Pulse Oximeter
Operation
43
– The “HomeCare mode” access protects the inexperienced user from modifying important
monitoring parameters. These parameters are invisible.
– The “Clinic mode” access allows users to set all the configuration parameters.
– The “Sleep Lab mode” will turn off all alarms during monitoring. This is useful for overnight
sleep studies.
CAUTION! Alarms cannot be detected!
If in Sleep lab mode an alarm condition occurs, the operator will not be alerted by
acoustical means.