RF Calibration Kit
24
Device Dependent Commands
The device-dependent commands used by the 4421 Power Meter are listed in
, organized by category.
Table 6 RS-232 Device Dependent Command Summary
F
orward
C
arrier Wave
(FC)
R
eflected
C
arrier Wave
(RC)
R
ange
(RYY)
Terminators
(YT)
P
refixes
(PY)
Category
Command
Description
Measurement
FC
Forward carrier wave
RC
Reflected carrier wave
Range
RYY
Auto range on
1
1 dummy command, the setting they control has only one option.
Terminators
YT
Two terminators: CR, LF
Prefixes
PY
Prefix YES
Triggers
T1
One shot on ENT
T5
One shot on measurement command
Function
Selects forward or reflected RF power measurement mode.
Remarks
Measurement results are returned in Watts.
Function
Sets the 4421A to auto range, legacy command to provide
compatibility with previous 4421 units.
Remarks
Auto is the only range setting available.
Function
Selects the characters that follow the end of a data string. YT
indicates for two terminators; a carriage return (CR) and a line
feed (LF).
Remarks
Many controllers use the terminator sequence to recognize the
end of an input string. Using incorrect terminators can lock the
bus.
Function
Turns the prefix mode on.
Remarks
Prefixes are sent over the bus with the measurement, and
indicate the status of the current measurement (see
for examples):
“FC”,”RC” indicates the measurement type.
“U” indicates underflow; the value sent is “.000”.
“O” indicates overflow; the value sent is “199.9”.
“N” indicates normal; the value sent is a normal on-
scale reading.
“4421” indicates the Bird model number.