RF Calibration Kit
20
4421A SCPI Commands
Common Commands
*IDN?
Description —
Identification Query
Parameter Type —
<arbitrary_ascii>
Parameter Minimum —
Parameter Maximum —
Parameter Default —
Parameter Values —
Parameter Units —
Response —
Returns device identification in the following format:
BIRD-TECHNOLOGIES,{model},{serial number},{firmware revision}
Measure Commands
MEASure[:SCALar][:POWer][:FORWard]:AVERage? [1 or 2]
Description —
initiate, and retrieve a forward average power measurment. The sensor channel to read (1 or 2) is
omitted, channel 1 is returned.
Parameter Type —
<NR1>
Parameter Minimum —
1
Parameter Maximum —
2
Parameter Default —
1
Parameter Values —
Parameter Units —
Response —
Returns the forward average power in W from the specified sensor channel. Parameter Type: <NR2>,
Parameter Units: W.
MEASure[:SCALar][:POWer]:REFLected:AVERage? [1 or 2]
Description —
initiate, and retrieve a reflected average power measurement.
Parameter Type —
<NR1>
Parameter Minimum —
1
Parameter Maximum —
2
Parameter Default —
1
Parameter Values —
Parameter Units —
Response —
Returns the Reflected average power in W from the specified sensor channel. Parameter Type:
<NR2>, Parameter Units: W.