49
Technical Data
Battery Data
Battery type characteristics
The following data is entered by the manufacturer:
Power consumption
The implant has the following power consumption:
Average service time
Average service times are precalculated using the battery manufacturer's technical
specifications, a basic rate of 60 ppm and the setting of different pulse amplitudes
and lead impedances.
Service times DR(-T)
For dual-chamber implants, the following times (in years) result:
Manufacturer
GREATBATCH, INC.
Clarence, NY 14031, USA
LITRONIK GmbH
01796 Pirna, Germany
Battery type
GB 8431
GB 2596
LiS 3150
LiS 3150M
System
LiJ
Ag/SVO/CFx
QMR
®
LiJ
LiMn0
2
Implant
DR
SR
DR-T
SR-T
DR
SR
DR-T
SR-T
Battery voltage
at BOS
2.8 V
3.0 V
2.8 V
3.1 V
Open-circuit volt-
age
2.8 V
3.0 V
2.8 V
3.1 V
Nominal capacity
1.3 Ah
1.3 Ah
1.3 Ah
1.2 Ah
Usable capacity
until EOS
1.2 Ah
1.1 Ah
1.2 Ah
1.0 Ah
Power consumption
DR(-T)
SR(-T)
BOS, inhibited
6
μ
A
6
μ
A
BOS, 100% pacing
13
μ
A
9
μ
A
Amplitude Impedance
[Ohms]
Pacing
10%
50%
100%
DR-T
DR
DR-T
DR
DR-T
DR
1.5 V
500
19.3
17.8
17.2
15.8
15.0
13.9
1000
19.7
18.0
18.3
16.9
17.0
15.7
2.5 V
500
17.8
16.6
13.6
12.8
9.4
10.0
1000
18.8
17.4
15.3
14.9
12.5
12.7
3.0 V
500
17.3
15.8
11.8
10.9
8.5
7.8
1000
18.5
16.9
14.6
13.4
11.6
10.7
3.5 V
500
16.5
15.1
10.4
9.7
7.2
6.6
1000
18.0
16.5
13.4
12.4
10.2
9.4
5.0 V
500
12.4
12.1
5.4
6.2
3.2
3.8
1000
15.2
14.6
8.3
9.2
5.3
6.3