After Implantation 5365353-ATechnical manual for the implantEvia DR-T, DR, SR-T, SR
Follow-up
Follow-up intervals
Follow-ups must be performed at regular agreed intervals.
• Follow-ups with the programmer should take place in intervals between
6 to 12 months considering the expected service life of the implant.
Follow-up with
the programmer
Proceed as follows:
1
Record and evaluate the external ECG.
2
Check the pacing function.
3
Interrogate the implant.
4
Evaluate the status and automatically measured follow-up data.
5
Possibly evaluate statistics and Holter/IEGM recording.
6
Manually perform standard tests if necessary.
7
Possibly customize program functions and parameters.
8
Transmit the program permanently to the implant.
9
Print and document follow-up data (print report).
10
Finish the follow-up for this patient.