39
Parameters
Lead configuration
IEGM recordings
Rates for statistics
Parameter
Range of values
Standard
Pacing polarity A
Unipolar; bipolar
Unipolar
Pacing polarity V
Unipolar; bipolar
Unipolar
Sensing polarity A
Unipolar; bipolar
Unipolar
Sensing polarity V
Unipolar; bipolar
Unipolar
Parameter
Range of values
IEGM recordings
20 (quantity); each max. 10 s
Types of IEGM recordings
High atrial rate (HAR)
Mode switching (MSW)
High ventricular rate (HVR)
Patient triggered (Pt.)
IEGM recording prior to event 0; 25; 50; 75; 100%
IEGM signal
Filtered; unfiltered
Parameter
Range of values
Standard
High atrial rate (HAR)
100 ... (5) ... 250 bpm
600; 572 ... 245; 240 ms
200 bpm
300 ms
High ventricular rate (HVR)
150 ... (5) ... 200 bpm
400; 378 ... 308; 300 ms
180 bpm
333 ms
HVR counter
4; 8; 12; 16
8