CardioMessenger-S TLine
Appendix
41
a)
Field strengths from fixed transmitters, such as base stations for radio (cellular(/cordless) telephones and land mobile radios, amateur
radio,
AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the
measured field strength in the location in which the CardioMessenger is used exceeds the applicable RF compliance level above, the
CardioMessenger should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary,
such as re-orienting or relocating the CardioMessenger.
b) Above the frequency range of 150KHz to 80MHz, ensure that field strengths are less than 3V/m.
CM-S_TLine.book Page 41 Wednesday, December 21, 2011 4:58 PM
Summary of Contents for CardioMessenger-S TLine
Page 4: ...2 CM S_TLine book Page 2 Wednesday December 21 2011 4 58 PM ...
Page 34: ...Precautionary measures 32 CM S_TLine book Page 32 Wednesday December 21 2011 4 58 PM ...
Page 50: ...CM S_TLine book Page 48 Wednesday December 21 2011 4 58 PM ...
Page 51: ...371017 B_GA_CardioMessenger S_TLine_en US_Cover indd 3 20 12 2011 19 08 26 ...