5. Reconnect it.
The CardioMessenger restarts and performs
the self-test. It checks the connection to the
BIOTRONIK Service Center. The operating
light switches to green once the connection
test is sucessful. The CardioMessenger is now
ready for use.
Note
The connection check can take up to
15 minutes.
Note
The CardioMessenger is designed to
operate with standard analog phone lines
providing Plain Old Telephone Service
(POTS). Compatibility with other phone lines
or services (e.g., Voice over IP (VoIP) via
analog or digital lines, cable TV phone lines
or ISDN) is not guaranteed. In these cases
compatibility testing at the patient’s home is
recommended. Proper set up of the system
at the patient’s home may require
installation of adapters and/or analog
phone lines. Please contact your telephone
service provider for information on the type
of phone line that you have.
If the operating light remains continuously
illuminated yellow, contact your cardiologist or
physician.