5. Technical specification

BS-3047 Series

Additional Objectives

0.5X

0.75X

1X(built-in)

1.5X

2X

Working distance (mm)

184mm

108mm

40mm

10X Eyepiece

/

φ

23mm

magnification

3.4

23.5

5.1

35.3

6.8

47

10.2

70.5

13.6

94

Field dia(mm)

67.65

9.75

45.1

6.52

33.82

4.89

25.55

3.26

16.91

2.45

15X Eyepiece

/

φ

17mm

magnification

5.1

35.3

7.65

52.9

10.2

70.5

15.3

105.8

20.4

141

Field dia(mm)

50.0

7.23

33.33

4.82

25.0

3.62

16.67

2.41

12.5

1.81

20X Eyepiece

/

φ

13mm

magnification

6.8

47

10.2

70.5

13.6

94

20.4

141

27.2

188

Field dia(mm)

38.24

5.53

25.49

3.69

19.12

2.77

12.75

1.84

9.56

1.38

BS-3048 Series

Additional Objectives

0.5X

0.75X

1X(Built-in)

1.5X

2X

Working distance (mm)

184mm

108mm

40mm

10X Eyepiece

/

φ

24mm

magnification

3.3

25.5

5.0

38.3

6.6

51

9.9

76.5

13.2

102

Field dia(mm)

72.73

9.41

48.48

6.27

36.36

4.71

24.24

3.14

18.18

2.35

15X Eyepiece

/

φ

17mm

magnification

5.0

38.3

7.4

57.4

9.9

76.5

14.9

114.8

19.8

153

Field dia(mm)

51.52

6.67

34.34

4.44

25.76

3.33

17.17

2.22

12.88

1.67

20X Eyepiece

/

φ

13mm

magnification

6.6

51

9.9

76.5

13.2

102

19.8

153

26.4

204

Field dia(mm)

39.39

5.10

26.26

3.40

19.70

2.55

13.13

1.70

9.85

1.27

* When use 0.5X auxiliary objective, the focus arm should be changed to upper position of the square rail pillar
stand.

No matter the zoom magnification, the working distance of same objective will not change.

Use the following formula to calculate Total magnification and field of view

Total magnification = Zoom Magnification × Eyepiece Magnification × Auxiliary Objective Magnification

FOV=Eyepiece FOV / (Zoom Magnification× Auxiliary Objective Magnification*)
* Without Auxiliary Objective, this value is 1

Summary of Contents for BS-3047 Series

Page 1: ...reo microscopes To insure safety and obtain optimum performance and familiarize yourself fully with the use of this microscope We recommend that you read the manual thoroughly before operating the microscope Attain this manual instruction in an easily accessible place near the microscope for the further reference BioTools Inc ...

Page 2: ...cing the bulb The correct bulb must be used as the same specification of the bulb in the microscope Use of other bulbs may damage the equipment Before replacing the lamp must turn off power switch and unplug it in order to avoid electric shock and damage to equipment When replacing the lamp be careful not to pollution bulb The bulb shouldn t have dust fingerprints oil etc The power supply must be ...

Page 3: ...1 Stand 6 3 2 Main body 6 4 Outfit 10 5 Technical specification 11 6 Installation 12 6 1 Installation diagram 12 6 2 Installation Details 13 7 Optional accessory 15 7 1 LED Ring light or Cold light source with Ring Optical Fiber 15 7 2 Polarizing unit 15 8 Troubleshooting 16 9 Maintenance and clean 16 ...

Page 4: ...BestScope International Limited 3 1 Parts Name ...

Page 5: ...BestScope International Limited 4 ...

Page 6: ... if correctly inserted into tube 3 Prepare for light source 2 2 Procedure 1 Put sample on stage 2 Adjust interpupillary distance 3 Adjust diopter 4 Adjust Zoom knob to minimum level rotate the focusing knob for coarse focus 5 Rotate the zoom knob to the required magnification then focus with focusing knob ...

Page 7: ...d light will be bright 3 Rotate down illumination wheel Fig 2 to adjust brightness 4 Rotate the top illumination for adjusting the light position 3 2 Main body 3 2 1 Interpupillary Distance Adjustment Fig 3 Holding binocular eyepiece tube adjusting eyepiece field until left right field overlap totally The noise is normal when rotating the eyepiece tube to adjusting field 3 2 2 Dioper adjustment Fi...

Page 8: ...ange from 5 to 5 or over 5 to 5 Whether over 5 or below 5 both can confirm by eyepiece length 3 2 3 Usage of zoom magnification limiter Fig 7 By using limiter indicate magnification value and zoom knob spacing ring Magnification can be limited in the required range Right zoom knob is used to limit the high magnification left zoom knob is used to limit the low magnification 1 Use hexagon wrench to ...

Page 9: ...onal objective Fig 9 Because 0 5X additional objective has long working distance bracket must adjust to higher position 1 With 5mm allen wrench to remove bracket screw up illumination lamp should be moved from the stand 2 Moving the bracket to higher position and fasten the screw Screwing additional objective into the hole at the bottom of the microscope 3 2 6 Storing for allen wrench Fig 10 Cover...

Page 10: ...etween image and video camera monitor for better image output after switching light path 1 Adjust eyepiece diopter to focus Fig 6 7 2 Rotate the switching knob to OUT set minimum magnification 3 Hold camera rotate adjustable knob adjust parfocality to focus for monitor image Jointing one 0 5X C mount with the bottom of the camera Monitor image rotation For TV light path The monitor image will rota...

Page 11: ...Eyepiece Φ24mm 20XEyepiece Φ12mm 0 5X 2X Objective Eyeguard Focusing unit No light Reflecting 3W LED Reflecting 5W LED Transmitting 3W LED Black white Plate Glass Plate Arm stand Pillar stand Base 330 300mm Polarizing Unit Fine focusing unit C mount 0 5X C mount 1X Adapter for digital photo camera Two dimensional travelling table Single pole Universal Stand Note In Table Is Standard Outfits Is Opt...

Page 12: ...ce φ24mm magnification 3 3 25 5 5 0 38 3 6 6 51 9 9 76 5 13 2 102 Field dia mm 72 73 9 41 48 48 6 27 36 36 4 71 24 24 3 14 18 18 2 35 15X Eyepiece φ17mm magnification 5 0 38 3 7 4 57 4 9 9 76 5 14 9 114 8 19 8 153 Field dia mm 51 52 6 67 34 34 4 44 25 76 3 33 17 17 2 22 12 88 1 67 20X Eyepiece φ13mm magnification 6 6 51 9 9 76 5 13 2 102 19 8 153 26 4 204 Field dia mm 39 39 5 10 26 26 3 40 19 70 2...

Page 13: ...ope International Limited 12 6 Installation 6 1 Installation diagram Note Number is installation sequence To ensure that all parts has no dust and dirt Note any part not scratch rub or touch the glass surface ...

Page 14: ...n body 2 Fasten the Body Locking screw You can rotate the main body if making observation from one side of focusing knob 6 2 2 Eyepiece Installation Fig 14 1 Turn eyepiece to 0 then insert into the eyepiece tube line is up if choose ordinary eyepiece no such step 2 Fasten the eyepiece locking screw with socket head wrench Fig 14 Fig 13 ...

Page 15: ...alled According to different sample to chose white plate cover or black plate cover 2 Please remove the plate directly with finger following the diagram 6 2 4 Camera installation Only apply to trinocular body Fig 16 1 Connect C mount and adapter ring 2 Screw in the combination adapter to the camera 3 Loosen the locking screw insert the camera and lock again 4 Connect the data cable from monitor or...

Page 16: ...wer after installing the illumination 2 Use light dimmer to change the illumination brightness 7 2 Polarizing unit 7 2 1 Installation 1 Put the analyzer in the bottom of the microscope head 2 Put the polarizing film in the stage hole 3 Install the glass plate Fig 15 7 2 2 Usage The polarizing unit must be used with transmitting illumination unit Focusing for samples rotate analyzer for best illumi...

Page 17: ...is out of focus Incorrect focusing adjustment for camera Focusing again 9 9 Maintenance and clean 9 1 If the lens surface has dust please clear with a blow ball If the lens surface with fingerprints or oil stain please clean gently with absorbent cotton or lens paper dipped in a little ethanol ether mixture ratio 1 4 Absolute alcohol is an inflammable material please turn off the power and don t c...

Reviews: