Service & Maintenance
Symptom
Possible Cause
Resolution
Display Code: S31
2 beeps when pressing any key.
2% life remaining notification for
Reverse Osmosis module
Replace the notified part. Contact your
dealer to arrange service ASAP.
Replace the notified part. Contact
your dealer to arrange service
IMMEDIATELY.
2% life remaining notification for
Deionization module
Replace the notified part. Contact your
dealer to arrange service ASAP.
2% life remaining notification for
Off-gas Destructor
Replace the notified part. Contact your
dealer to arrange service ASAP.
Replace the notified part. Contact your
dealer to arrange service ASAP.
Replace the notified part. Contact your
dealer to arrange service ASAP.
Replace the notified part. Contact your
dealer to arrange service ASAP.
2% life remaining notification for
System Cooling Fan.
2% life remaining notification for
Source Water Hose.
2% life remaining notification for
iEOG Cell.
Service due and life ending
notification for Reverse Osmosis
module
Service due and life ending
notification for Deionization
module
Replace the notified part. Contact
your dealer to arrange service
IMMEDIATELY.
Service due and life ending
notification for Off-gas Destructor
Replace the notified part. Contact
your dealer to arrange service
IMMEDIATELY.
Replace the notified part. Contact
your dealer to arrange service
IMMEDIATELY.
Service due and life ending
notification for System Cooling Fan.
Service due and life ending
notification for Source Water Hose.
Replace the notified part. Contact
your dealer to arrange service
IMMEDIATELY.
Contact your dealer to arrange
service IMMEDIATELY.
Notification for iEOG Cell.
Display Code: S31
All functions suspended.
10 sec alarming, followed by 5 beeps
when pressing any key.
Display Code: S32
All functions suspended.
10 sec alarming, followed by 5 beeps
when pressing any key.
Display Code: S34
All functions suspended.
10 sec alarming, followed by 5 beeps
when pressing any key.
Display Code: S38
All functions suspended.
10 sec alarming, followed by 5 beeps
when pressing any key.
Display Code: S39
All functions suspended.
10 sec alarming, followed by 5 beeps
when pressing any key.
Display Code: S71
All functions suspended.
10 sec alarming, followed by 5 beeps
when pressing any key.
Display Code: S38
2 beeps when pressing any key.
Display Code: S39
2 beeps when pressing any key.
Display Code: S71
2 beeps when pressing any key.
Display Code: S32
2 beeps when pressing any key.
Display Code: S34
2 beeps when pressing any key.
31