TurboVap
®
96 Dual User Manual | © Biotage 2021
Evaporate
Change a Setting in Manual or Time Mode
To change the settings in Manual or Time mode, proceed as
follows:
1. Press the button for the setting to be changed (gas flow, gas
temperature, plate temperature or plate height), see Figure
20 and Figure 21.
2. Press the
Up and Down
arrows in the appearing dialog
to change the setting step by step. Alternatively, use the
numeric keypad to enter numbers, see Figure 22.
1
Change a setting by pressing the Up and Down arrows
2
Change a setting by entering numbers
Figure 22.
The Change a setting dialog. In this figure the setting to be
changed is the gas flow for the left cabinet.
Split to Left and Right (Single Mode)
The Split to left and right mode enables the left and right
cabinets to operate independently. The left cabinet can, for
example, operate a Manual mode run and at the same time the
right cabinet can operate a Time mode run. The two cabinets
can also operate individually with predefined methods.
For instructions regarding method configuration, see "Method
Mode" on page 14.
Join Left and Right (Dual Mode)
In the Join left and right mode, the left and right cabinets
operate the same evaporation run with the same parameters,
see Figure 23.
Figure 23.
Temp Step Method 96 run in Join left and right mode (Dual
mode).
Note:
When using Dual mode, the manifolds in the left and right
cabinet must be of the same type, see Figure 24. To configure
the manifold settings, see "Common Settings" on page 8.
The Idle Heating settings can differ between the left and right
cabinets. The system will wait for all system heaters with
Idle Heating settings set to
On
or
Auto
to reach their target
temperatures before starting an evaporation run.
Figure 24.
In order to select the Join left and right mode, the left and the
right manifolds need to be of the same type. In this figure the manifold
settings are unsynchronized (96 and 24, respectively).
Method Mode
It is possible to save up to 10 different evaporation methods per
manifold configuration. The methods can be set up to evaporate
using a ramp gradient of gas flow, gas temperature and plate
height with a maximum of four time segments. The left side and
the right side can be set up individually, running two separate
methods at the same time. The left and right side can also
operate as a dual unit, performing the same evaporation run
with the same evaporation parameters.