28
C. Standing Posture
(1) It is recommended that the examinee standing posture for about 10~15 minutes before
the test, so that body water may be dispersed evenly inside the body.
(2) Make sure bare feet do not touch the floor. Use a mat that does not conduct electricity.
(3) Make sure arms do not touch the trunk part of body. Spread them naturally to a 15
degree angle away from trunk.
(4) Arms should be posed lowered naturally.
(5) Make sure thighs do not touch each other, and spread legs to shoulder width.
Attention
• It is recommended that the examinee standing posture for about 10~15 minutes.
• Make sure bare feet do not touch the floor. Use a mat that does not conduct electricity.
• Hang the electrode cables to not let them dragging on the floor.
• For a examinee with curvy feet the foot electrodes might not attach properly when using
the Touch Type electrodes. Make sure to attach it as best as possible.
• If examinees have dry hands and feet, wipe the equipment with an electrolyte tissue
before test.
• Due to the weight of the cable, it is easy for the Adhesive Type electrodes to fall off
during the test. Please be careful and make sure that the electrodes are attached
properly.
Summary of Contents for inbody s10
Page 18: ...7 When you use the touch type electrodes When you use the adhesive type electrodes ...
Page 25: ...14 Thermal printer cables are connected in the same manner as shown in the picture ...
Page 52: ...40 4 ICW ICW ECW History 5 TBW FFM History 6 Weight Muscle PBF History ...
Page 62: ...50 Chapter3 Setup Establishment 1 Setup 2 Setup Menu 3 DATABASE ...
Page 77: ...Chapter5 Consumables 1 Consumables 2 Options ...
Page 84: ...69 IB Battery 110 connection diagram When use InBody S10 bag When use InBody S10 cart ...
Page 85: ...70 Appendix 1 More about the InBody S10 2 Specifications 3 Customer Service Information ...