7
5.
Perform a Verification Test with
OD Plate Verification Software
5.1.
For the verification, use the OD Plate (Catalogue number BS-
050108-AK)
5.2.
Plug the flash drive provided with the plate in the computer.
5.3.
Open the
OD Plate verification software.exe
5.4.
Choose the verification plate from the list. The 4 digits number stick-
ered on the side of the
plate is the plate’s ID number.
5.5.
Choose the channels you need to verify. If you have additional chan-
nels, choose them. The wavelength of
Channel 1/2/3/4
are written
body of the unit.
5.6.
Set the number of measurements. Recommended is 20, minimum is
10.
5.7.
Enter the name of the QC operator, Head of QC and QC company in
the empty fields.
5.8.
Press
Start
.
5.9.
When finished, press
Results
button.
5.10.
Test type should be
Verif
. Other test types are for manufacturers
use only.