background image

 

10 

7. 

Troubleshooting 

7.1. 

The accuracy or precision test have failed

 

In case of failed tests, the reason may be in faulty or unclean instru-
ment  or  OD  Plate.  Generally,  if  one  or  more  of  the  precision  tests 
have FAILED, the reason lies in the instrument or OD plate, if it was 
assembled. If one or more of the accuracy tests have FAILED, the 
reason may lie in the instrument or in the OD Plate. Here is the list of 
to do things: 

 

Check  that  the  OD  Plate  has  been  placed  correctly  onto  the  plate 
carriage. The A1 markings on the instruments should coincide and 
the OD plate should lie flat. 

 

Check that the glasses are clean. 

 

If  the  fault  does  not  disappear, contact the technical service  repre-
sentative. 

7.2. 

The device cannot connect to the computer

 

Install the driver 

by opening “Drivers/Driver.exe” on the provided USB 

flash drive and restarting your devices/software/computer after. 

7.3. 

The program cannot close, says that the experiment is still go-
ing, but I stopped it

 

Try pressing the 

Play

 button on the toolbar (to start the experiment), 

and click on the 

Stop

 button, wait 5 seconds and then try to close it. 

If  this  does  not  work,  open  Task  Manager  (Ctrl  + Shift  + Esc)  and 
close all processes “OD plate verification software.exe” 

7.4. 

Device does not respond to the program

 

Try  turning the  device  on  or  off.  If  necessary,  try  unplugging  it  and 
reconnecting. 

 

 

Summary of Contents for OD Plate

Page 1: ......

Page 2: ...2...

Page 3: ...oduction 5 3 Specifications 6 4 Complete Set 6 5 Perform a Verification Test with OD Plate Verification Software 7 6 Interpretation of the Measurement Results 9 7 Troubleshooting 10 8 Storage 13 9 Cle...

Page 4: ...see package la bel Before using any cleaning or decontamination methods except those recommended by the manufacturer check with the manufacturer that the proposed method will not damage the equipment...

Page 5: ...14 450 480 492 515 540 550 560 568 580 594 620 632 650 690 700 Optionally the plate can be verified at any following Centre Wave lengths CWL 1 nm 400 405 414 420 422 430 436 440 442 445 450 455 458 46...

Page 6: ...1 0 2 0 3 0 4 0 Abs 3 2 Available calibration wavelength range 52 in range 400 to 700 nm 1 nm 3 3 Plate Dimensions 128 x 86 x 12 mm 3 4 Plate Weight 0 2 kg 3 5 Plate Design 4 Complete Set OD Plate USB...

Page 7: ...gits number stick ered on the side of the plate is the plate s ID number 5 5 Choose the channels you need to verify If you have additional chan nels choose them The wavelength of Channel 1 2 3 4 are w...

Page 8: ...order to export the verifi cation certificate Also available is the csv export 5 13 Print the report sign it and leave with the customer 5 14 Save the report and send it to the manufacturer qc biosan...

Page 9: ...st values meets the specifications Accuracy Criteria 0 2 3 OD 0 5 0 010 OD typical 2 3 3 OD 1 0 010 OD typical 3 4 3 OD 2 0 010 OD typical 4 3 5 OD 15 0 010 OD typical 6 2 Precision Precision here is...

Page 10: ...uld lie flat Check that the glasses are clean If the fault does not disappear contact the technical service repre sentative 7 2 The device cannot connect to the computer Install the driver by opening...

Page 11: ...ghts to the user who installs the program 7 5 1 If the previous step did not help try the following Go to Control Panel Device Manager Expand the Other devices line right click the USB Serial Port Cli...

Page 12: ...and install the driver on the computer to which the units are connected 7 5 3 If it does not help please try a solution described here https support microsoft com en us help 2654149 error usb device...

Page 13: ...seen in the apertures Is this OK The working aperture of the HiPo MPP 96 is only 2 mm in diameter so if you have received the plate with some small part of the stickers located inside the bigger apert...

Page 14: ...sunk screws from the sides Unscrew the facing to you bottom side We suggest taking a picture on a smartphone for the ease of reas sembly Carefully remove the glass for cleaning 9 4 Handle the glasses...

Page 15: ...s then used to calibrate the Reference Instru ment The glasses from the OD plate is then calibrated by measuring the absorbances of the glasses with the reference instrument The uncertainty of the cal...

Page 16: ...Edition 1 02 March of 2020...

Reviews: