Care and maintenance 

8.1 

Service

.  

8.1.1  If the unit is disabled (e.g., lamps not working, no reaction on key and switch presses) 

or requires maintenance, disconnect the unit from the mains and contact Biosan or 
your local Biosan representative. 

8.1.2  All maintenance and repair operations (except listed below) must be performed only 

by qualified and specially trained personnel. 

8.1.3  Operating integrity check. If the unit follow the procedure described in section 

Oper-

ation

, then no additional checks are required. 

 

 

 

 

 

Figure 4. 

Figure 5 

Figure 6 

8.2 

Fuse replacement

8.2.1  Fuse for the control box. Disconnect the unit from electric circuit. Remove power ca-

ble from its socket on the rear side of the unit. Open the fuse holder by pulling out the 
holder (fig. 4, compact models) or by screwing off the holder lid marked FU2 (fig. 6, 
model 

UVT-S-AR

). Check the fuse and replace if necessary, 

М

 3.15 A (type 

M

 - time 

lag: 

M

edium).  

8.2.2  Fuse for the mains outlets (for models with built-in mains outlets). Disconnect the unit 

from electric circuit. Open the fuse holder by screwing off the holder lid (fig. 5/2 or 
fig 6,  FU1).  Check  the  fuse  and  replace  if  necessary, 

М

  5.0  A  (type 

M

  -  time  lag: 

M

edium). 

8.3 

UV lamp replacement

. Average lifetime of UV lamps supplied is 9000 hrs. Replace-

ment is necessary after lamp stops functioning or at the end of manufacturer specified 
lifetime.  

 

Note. 

Open  UV  lamp  operation  can  be  checked  with  the  visible  light  lamp 
switched off (switch 

1

 is OFF). Use the indicator in the centre of recircu-

lator cover (fig. 1/3) to check operation of UV lamp inside the recirculator. 
If  indicator  lights from  inside while  switch 

2

  is  ON, then  the  UV lamp  is 

functioning.  

 

 

Summary of Contents for BS-040102-AAA

Page 1: ......

Page 2: ...services we would like to hear from you Please send all feedback to Manufacturer SIA Biosan Ratsupites iela 7 k 2 Riga LV 1067 Latvia Phone 371 674 261 37 Fax 371 674 281 01 www biosan lv Marketing m...

Page 3: ...recautions Caution Make sure you have fully read and understood present instructions before using the equipment Please pay special attention to sections marked by this symbol Caution The front protect...

Page 4: ...label Do not plug the unit into an ungrounded power socket and do not use an ungrounded extension lead Ensure that the power plug is easily accessible during use Disconnect the unit from the mains bef...

Page 5: ...ed to UV radiation It allows continuous treatment of the airflow with UV light without interrupting working process Air circulation at a short distance from the UV lamp combined with reflective surfac...

Page 6: ...w the instructions included on the pack aging 4 2 Complete set Package contents 4 2 1 Standard set DNA RNA UV cleaner box 1 pce Power cable UVC T AR only 1 pce Spare fuse for control block1 1 pce Spar...

Page 7: ...cable fig 5 1 is connecting the fuse box and the controller top 5 1 Connect the power plug to a grounded power socket Turn on the Power switch UVC T AR no switch the unit powers up automatically UVC T...

Page 8: ...e unit 5 2 5 If the set time of open UV light exposition is 0 00 pressing the Start key will make the unit operate continuously during 24 hrs or until the Stop Reset key is pressed 5 3 The box is read...

Page 9: ...ent and reserves the right to alter design and specifications of the equipment without additional notice Model UVC T AR Compact UVC T M AR Compact UVT B AR Compact UVT S AR Double Back panel PMMA Stee...

Page 10: ...Dimensions 720x535x555 mm 700x580x555 mm 1250 x 600 x 590 mm Weight net gross 2 23 33 kg 28 8 39 kg 31 2 42 kg 58 68 5 kg Laboratory table T 4 T 4L 6 1 Used materials PMMA polymethylmethacrylate acryl...

Page 11: ...AB V 5A7 1 3 x US type B BS 040107 AA7 V 5A04 1 3 x AU type I BS 040107 A04 V 5AI 1 3 x IL type H BS 040107 AIL Caution Overall consumed power of devices connected through internal mains sockets shoul...

Page 12: ...y pulling out the holder fig 4 compact models or by screwing off the holder lid marked FU2 fig 6 model UVT S AR Check the fuse and replace if necessary 3 15 A type M time lag Medium 8 2 2 Fuse for the...

Page 13: ...ing cleaners more than 20 or ammonia containing cleaners for acrylic glass Do not use abrasives The table below shows the interaction of acrylic glass with ethyl alcohol and other solutions Solution I...

Page 14: ...ocal distributor address To obtain the claim form visit Technical support page on our website at link below 10 5 Extended warranty For UVC T M AR UVT S AR the Premium class model one year of extended...

Page 15: ...ts for electrical equipment for measurement control and laboratory use General requirements EMC 2014 30 EU LVS EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requi...

Page 16: ...Biosan SIA Ratsupites 7 k 2 Riga LV 1067 Latvia Phone 371 67426137 Fax 371 67428101 http www biosan lv Edition 3 6 02 March of 2021...

Reviews: