BIOS Defender 500 Series User Manual Download Page 9

14

15

2.7 Measurements

Taking gas flow readings

To maintain the best possible accuracy and minimize thermal  
effects, Bios recommends fully charging your battery before 
taking measurements. If this is not possible, we recommend 
disconnecting your Defender from its AC power adapter/charger 
while taking flow measurements — or to run gas through your 
Defender for 10 minutes before starting the flow measurement.

First steps
Press the power button.

•   Press the On/Off button for 1 second to turn on your Defender.
•   When first turned on, your Defender displays an opening 

screen showing the product name, model number and flow 
range.

•   Press the On/Off button for 3 seconds to turn your  

Defender off.

Connect device to be calibrated to the appropriate Defender port. 
Use ¼ inch diameter tubing. 

•   Connect to outlet at top (suction fitting) when a device draws 

air (such as sampler). 

•   Connect to inlet at bottom (pressure fitting) when a device 

pushes air.

•   Do not cap the unused port on the Defender.
•   Choose the measurement type, Single, Burst, or Continuous, 

then press enter.

2.8 Single Measurement

Each time the ‘Enter’ button is pressed, one measurement will 
be made.  When each subsequent measurement is made, the 
current flow and average of all prior readings will be displayed.

2.9 Burst Measurement

This setting functions in the same manner as ‘SINGLE’,  but 
measurements continue automatically until the preset number of 
measurements has been made.  Operation then ceases, and the 
last reading and average are displayed.

Press ‘Enter’ again to begin another preset sequence.

2.10 Continuous Measurement

This setting functions in the same manner as ‘BURST’, but new 
sequences will automatically repeat until stopped by the user.

2.0

Summary of Contents for Defender 500 Series

Page 1: ...Defender 500 Series User Manual Driving a Higher Standard in Flow Measurement SM...

Page 2: ...ftware 3 1 SKC CalChek Interface 3 2 Defender Firmware Upgrades 3 3 Annual Maintenance and Calibration 4 0 Shipping 5 0 Storage 6 0 Defender Specifications 7 0 Default Settings 8 0 Limited Warranty 9...

Page 3: ...you have questions regarding its opera tion please contact Bios through our web site www biosint com or call us at 973 492 8400 to speak with a member of our professional customer service staff Your D...

Page 4: ...pressure devices 2 1 Battery Charging installing and monitoring your Defender battery Your Defender battery is charged at the factory but we recommend that you make sure it is fully charged before in...

Page 5: ...The Defender provides a menu of operational settings and com mands The four directional arrow buttons on the control panel allow you to navigate through the menu and select the desired settings for y...

Page 6: ...If you wish to incorporate a time delay between consecutive measurements set Time Between from 1 to 60 minutes Units Measure gas Flow in cubic centimeters milliliters liters or cubic feet all units a...

Page 7: ...SKC CalCheck Interface Magnification In the Defender 520 model this controls the amount of data on the display Select Zoom to view only flow measurements in larger font or select Detail to simultaneo...

Page 8: ...Bios supplied leak test cap Leave the other port uncapped Press Enter on the control panel while the unit is still inverted Return the unit upright The leak test will progress If Defender leak test i...

Page 9: ...er port Use inch diameter tubing Connect to outlet at top suction fitting when a device draws air such as sampler Connect to inlet at bottom pressure fitting when a device pushes air Do not cap the un...

Page 10: ...terface for use with Bios Optimizer Software or with the SKC CalCheck Interface Visit www biosint com to find more information about Optimizer software and for access to available Optimizer or Defende...

Page 11: ...ined to extremely close tolerances Various environ mental factors product wear drift of sensors or inadvertent damage may adversely affect your Defender s measurement accuracy or general performance F...

Page 12: ...er warranty Tips and guidelines for sending your Defender to Bios If you are sending your Defender for repair or evaluation rather than elective Recertification contact Bios for technical support or t...

Page 13: ...storage 6 0 7 0 7 0 Defender Specifications Technical data about your Defender Models 510L 520L from 5 500 mL min 510M 520M from 50 5 000 mL min 510H 520H from 300 30 000 mL min Measurements Volumetri...

Page 14: ...e PC and RS 232 serial connection Bios Optimizer installation CD supplied no restrictions apply RS 232 cable supplied for Defender 500 Series data port to PC RS 232 serial port connection PC Card opti...

Page 15: ...ration shall have no liability to repair or replace any Bios International Corporation product 1 That has been damaged following sale including but not limited to damage resulting from improper electr...

Page 16: ...der for pressure sources or to the suction port for ver ifying suction pumps The unused port should be at atmospheric pressure with any cap or plug removed If you are calibrating a gas that requires a...

Page 17: ...of the cell after a measurement this could be caused by A discharged battery not providing enough power to operate the internal valve properly Try charging the Defender Bright light shining into the...

Page 18: ......

Page 19: ...Measurement SM Bios International Corporation 10 Park Place Butler NJ USA 07405 Phone 973 492 8400 Toll Free 800 663 4977 Fax 973 492 8270 Email sales biosint com web www biosint com 2009 Bios Intern...

Reviews: