17
ATTACH FOOT MODULE
(If necessary)
Before installing the Foot Module, determine if a foot shim is necessary by turning
the ankle over and looking at the bottom of the foot structure. If “NO SHIM” is
printed proceed to Step 2 “INSTALL FOOT MODULE”. If there is no inscription on
the foot structure follow instructions in Step 1 “INSTALL FOOT SHIM”.
1. INSTALL FOOT SHIM
• Install onto threaded bosses of foot structure with
the flat side of foot shim facing away from foot
structure.
2. INSTALL FOOT MODULE
• Install the Foot Module by aligning the two holes
with the bosses located on the bottom side of the
foot structure. Ensure the toe end faces forward.
• Secure the Foot Module with two bolts and a .05”
washer on each bolt (provided with system kit).
• Using a wrench with a 1/2” hex head, torque the
bolts to 110 + 5 in-lbs
3. INSTALL FOOT COVER
• Pull the spectra sock over the Foot Module.
• Slide toe end of the Foot Module into the Foot
Cover. Next (using a shoe horn if necessary), fit
the heel into the Foot Cover and push down until
it pops into place. Confirm the Foot Module heel is
captured below the interior ridge at the base of the
Foot Cover.