15
ALIGNMENT
INITIAL ALIGNMENT - TRANSTIBIAL
1. ESTABLISH NEUTRAL POSITION
• Ensure Ankle Housing is in contact with the Hardstop
Cover.
• If the heel height of the shoe is greater than 3/8”
(9.5mm), extend the pylon to achieve a neutral
position.
2. ADJUST BATTERY HOLSTER TO AVOID INTERFERENCE
WITH SOCKET AND CONNECTORS
• Loosen Battery Hinge screws with a 2.5mm hex key
until the Battery Hinge can move freely.
3. ATTACH SOCKET TO THE ANKLE
• Attach socket to the Integrated Male Pyramid on the
Ankle.
• Do not exceed 5° of external rotation of the ankle
assembly.
• Reference line should bisect the socket, and fall at
the junction of the posterior and middle one third of
the foot.
4. POSITION BATTERY HOLSTER AS CLOSE TO THE
SOCKET AS POSSIBLE
• Loosen upper Battery Hinge screws to further
customize the Battery position.
• Hand-tighten all 8 Battery Hinge screws when Battery
Holster is in desired position.
• For the lower screws, be sure to tighten screws
within the scallop design.
• Check that the Battery will easily clear the socket
when removing from the Battery Holster.