51
Error: Not Supported Device
Solution
Use the cartridge barcode provided by BIONOTE.
Error: Optic Detail Error.
Error in M10 Optical Measurement Module.
Solution
Turn off the power of M10 Console and M10 Module and
check if the cable and gender are properly connected.
Error: Check the connected module information
Connect UI number and it must be identical to the number of
physically installed M10.
Solution
If the Connect number is displayed differently from the Total
number, turn off the power of M10 Console and M10 Module
and check if the cable and gender are properly connected.
Error: Network setting is not properly completed.
The setting of network related to the Network and LIS of
M10 Console is incorrect.
Solution
Check the network settings again, and contact the network
administrator.