45
CHAPTER 6
Cleaning and Maintenance
1. Analyzer Cleaning
To prevent malfunction of the analyzer, a test cartridge inlet should be free of moisture or dust
from the specimen. Clean the analyzer using lint-free cloth and appropriate cleaning solution
(ex: mild soapy water, 70% of ethanol or isopropyl alcohol etc.). When using the analyzer for
professional purposes (ex: operating room), it is recommended to clean it with a mixture of
1-propanol, 2-propanol, and glutaraldehyde (trade name “Bacillol plus”).
Do not use abrasive cloth or antiseptic solution as this may damage the
display screen. Always turn off the analyzer and remove the power cable
before cleaning.
2. Maintenance and Transportation
If there is any problem on the Vcheck M10 system, the Console automatically shows what went
wrong for each time the system is powered on.
• Be careful that test device insertion and analyzer internal parts should
not be contaminated by dust.
• Transport case is designed to store the supplied components and
protect the analyzer.
•
Store the analyzer at -20°C ~ 50°C / 0% ~ 93% RH.
Summary of Contents for Vcheck M10