24
When the connection is completed, you can use the analyzer.
• Re-installation is required for following situations.
- When the number of M10 Module has changed.
- When the position of M10 Module has changed.
- If the number of connected M10 Module and the number of M10 Module recognized on UI
are different. Please check the connection cables or gender.
• Re-installation Procedure
- Only the admin account is able to proceed this task.
- Click the install button on the Settings menu to proceed.
- During the process, the yellow LED light may flash.
- If the problem continues, check the connection between the console and modules. Turn off
all power at least for 5 seconds and reboot.