49
48
Performing a Quality Control Test
Performing a Quality Control Test
9. Tightly replace the cap on
RIGHTEST Max Control Solution bottle.
10. The control solution result will appear. Compare your Quality Control Test result
to the Control Solution Range printed on
RIGHTEST Max Blood Glucose Test
Strip vial label.
CLOSE
9
CAUTION
- Your Control Solution Test results will not be included in the average calculations, however,
they can still can be recalled and viewed. The Control Solution Test result will be shown
with the " " icon on the screen.
- The Control Solution Test should be conducted between 6 - 44°C (43 - 111°F), relative
humidity 10 - 90%.
- Before " " and " " appear, do not touch the control solution to the sample port on
the strip. RIGHTEST Blood Glucose Meter is performing an internal check. Touching the
control solution to the sample port before prompted will result in an error message:
" " and " " and be accompanied by beeps (if volume is turned on).
- Do not drip the control solution on to the sample port of the test strip directly. The reagent
on the strip might leak into the bottle of control solution and may cause the degeneration
of the control solution. This could contaminate the meter via the test strip port.
- Keep the test strip port clean and dry. Clean immediately if the test strip port is
stained or is overly exposed to moisture.
- Do not touch the tip of the control solution bottle. If the tip is touched, clean
with water.
6
OPEN
5
3
1
2
7
1. Take one
RIGHTEST Max Blood Glucose Test Strip
from the vial and
close the vial cap immediately.
2. Insert the strip into the test strip port of
RIGHTEST Max Blood
Glucose Meter
with the view window facing up.
3.
While the test strip icon is flashed on the display window, press
and hold the main button for at least 3 seconds until the " "
symbol appears.
4. You will see "
" icon and " " icon blinking on the screen,
prompting you to apply the corresponding level of
RIGHTEST Max
Control Solution
.
5. Shake the bottle of
RIGHTEST Max Control Solution
well before
opening the cap. Place the cap on a flat surface.
6. Place a drop of control solution onto the top of the cap.
7. Gently touch the sample port of the strip with the control solution from
the top of the cap.
8. The screen will display the count time starting from 5 (you will hear a
beep if the volume is turned on).
8
Summary of Contents for RIGHTEST Max
Page 2: ......