View Window
This window is yellow before applying
blood sample. It gradually turns red
when filled with blood.
Hand Bar
Grip here to insert
test strip into meter.
Electrode Contacts
Sensing signal output terminals.
Noble Metal Electrodes
Electrochemical sensor
Indication Symbol
Insert strip with indication symbol
up and toward meter.
The
Rightest™
Blood Glucose Meter GM260 can only be used with the
Rightest™
Blood Glucose
Test Strips GS260.
The application of other Test Strips or control solutions can lead to incorrect results.
TM
The
Rightest
Blood Glucose Test Strip GS260
Sample Entry
µL
Apply a drop of blood or control solution here.
The test request is only 0.75
of blood.
TM
The
Rightest
Blood Glucose Test Strip GS260
PRECAUTION
TM
-
Re-cap the
Rightest
Blood Glucose Test Strip GS260 vial immediately after taking out a
test strip.
TM
-
Do not reuse
Rightest
Blood Glucose Test Strip GS260.
TM
-
Do not use expired
Rightest
Blood Glucose Test Strip GS260.
TM
-
Record the date of opening a new
Rightest
Blood Glucose Test Strip GS260 vial for the
first time. Discard the vial of test strips after 3 month from opening.
TM
-
Store the
Rightest
Blood Glucose Test Strip GS260 in a cool (39-86
℉
or 4-30
℃
) and
dry location (< 90% relative humidity). Do not expose to direct sunlight or heat.
TM
-
For detailed information, please refer to the
Rightest
Blood Glucose Test Strip GS260
Package Insert.
TM
- If the
Rightest
Blood Glucose Meter GM260 and Test Strips GS260 are exposed to a
high temperature difference, please wait 30 minutes before measurement.