39
38
Performing a Quality Control Test
CLOSE
9
Recalling Test Result and Average
2.
To finish reviewing memory tests, press Main button again, and you will enter the Average screen.
By pressing the Right button, the display shows "AVG" on the upper right corner of the display and
the average value for the actual day. By sequentially pressing the Right or Left button you will get
to the 1-day, 14-day, 30-day and 90-day average of your blood glucose values. You will see the
number of the calculated days on the lower left corner and the number of the calculated readings
on the lower right corner.
1. Press the Main button to switch from the Time mode to the Memory mode.
First
you will see the ”MEM” symbol on the up per left corner of the display. When
you press the right button, the latest result will be displayed. By pressing the
right button sequentially former test results will be shown in historical order. You
will see the sequence number on the lower right corner and the year on the
lower left corner of the display followed by date and time of the measurement.
Note: The Right button is for reviewing the tests with sequence number increasing, and the Left
button is for reviewing tests with sequence number decreasing. The sequence no “1” is the latest
result while sequence no “500” is the oldest test result.
TM
9. Tightly replace the cap on the
Rightest
Control Solution bottle.
10. The Control Solution result appears. Compare your Quality Control Test result to
TM
the Control Solution Range printed on the
Rightest
Blood Glucose Test Strip
vial label.
GS260
PRECAUTION
- Your Control Solution results will not be calculated for average reading but still can be
recalled. The Control Solution Test result will be shown with " " symbol on the screen.
- Our suggestion environment for Control Solution Test is 59-104°F
°C .
TM
- Before " " and " " appears, please don't touch the
Rightest
Control Solution to the
TM
sample entry on strip because the
Rightest
Blood Glucose Meter GM260 is still in an
internal check. If you do so, the meter will show " " and " ". Accompanied by beeps
(if volume is turned on).
TM
- Don't drip the
Rightest
Control Solution to sample entry of the strip directly.
TM
The reagent on strip might be sucked into the bottle of
Rightest
Control
TM
Solution and might cause the degeneration of
Rightest
Control Solution.
Doing this might contaminate the meter via the test strip port as well.
TM
- Don't touch the tip of the
Rightest
Control Solution bottle. If you have
touched it, please clean up with water.
(15-40
)
The
GM260 is able to store 500 test results with time and date
automatically. If your meter has stored 500 results, which is the maximum memory of the meter, the
newest test result will replace the oldest one. To recall your test memory, start the meter without test
strip inserted.
TM
Rightest
Blood Glucose Meter
Summary of Contents for rightest GM260