Solution for precaution and warning
49
BFM-900L
FETAL MONITOR
(Ver 1.1)
48
Setting by Menu Function
6.5 Default Setting
Every basic parameter except for time on the device is initialized.
Step 1)
Press the Print key( ) + UC REF key( ) at once in the turning the device off condition.
Step 2)
Turn on the device with pressing the above keys. All basic parameter except for time is
initialized. If initialized, the screen will be like below.
Step 3)
Then turn on the device without pressing the keys. All parameter is initialized.
CHAPTER 7
Solution for precaution and warning
7.1 Solution for Alarm sound and Message
When the values go over or below the Alarm range over 10 seconds, the beep will sound. The
color of alarm LED lamp will be changed green to red. The FHR will be also twinkled.
7.2 Solution for Information sound and Message
①
The beep sounds once like “ding dong” as information sound when turn on the device.
If you want to turn off the alarm, press the SILENCE key.
Error Code
Code
E1
E2
E3
E4
E5
Description
No paper or paper out during printing
Disconnected DOP1 from DOP1 connection part during using
Disconnected DOP2 from DOP2 connection part during using
If “E2” error occurs, it will not be displayed
Disconnected UC from UC connection part during using
If “E2” and “E3” error occurs, it will not be displayed
Low Battery Low
If “E2” ~ “E4” error occurs, it will not be displayed.