Electrocardiograph (Ver 1.0)
8
Operational Do
’
s and Don
’
ts
Precautions for Electric Safety
9
BCM-600
◈
The product may not work well when it is kept or operated beyond the temperature and
humidity specified.
◈
If you want to turn off the BCM-600 while you’re operating, please use the power switch
where is the front of the product.
◈
When you keep or operate the product in large temperature difference, product
malfunctioning can be happened because of humidity of battery pack.
Please operate product after keep it room temperature for three hours.
◈
Leakage, heating, ignition or rupture of the battery may cause fire or injury.
Do not use batteries beyond those specified in this manual.
Do not short-circuit the battery and do not heat, disassemble, or dispose of it in a fire.
Do not mix using the used battery with new batteries , different kind of batteries, other
company batteries
Do not insert batteries with (+) and (-) ends reversed.
◈
When the product is not used for a long time, separate the battery from the product.
◈
To ensure the lifecycle and safety of battery, please discharge the battery and then charge
perfectly again every six months.
Also time-worn battery exceed 24 months should be replaced.
◈
When the product has been moved, make sure to turn it off and arrange the accessories in
order. Injured cords or cables may cause fire or electric shock.
Precautions for Electric Safety
To prevent electric noise during use, the product should be installed apart from dynamo, X-
ray equipment, broadcast equipment or portable cables. An inaccurate result may occur when
these equipment are placed near the product. BCM-600 requires an independent power circuit
and a stable earth connection. Sharing power with other electric equipment may result in
inexact results.
BCM-600 is classified as follows :
◈
Type 1 for electric shock, CF grade
◈
Please do not use this product near inflammable anesthetics and diluents.
◈
Calssification is Class IIb
◈
The noise level is “B” Class according to the IEC/EN 60601-1 (Safety of Electric Medical
Product), and the noise redemption is “B” Level according to the IEC/EN 60601-1-2
(Electromagnetic Compatibility Requirements)
◈
Protection grade against water is “IPX 0”.
Please check the following points before use.
Is the product supplied with the appropriate power? (100~240VAC)
Are all connecting parts (power cables or the product) properly connected?
Is the product completely grounded? (If not, noise may occur.)
Before turning the power on, is the accessory for measuring the target parameter well connected
to the product?