Electrocardiograph (Ver 1.0)

8

Operational Do

s and Don

ts

Precautions for Electric Safety  

9

BCM-600

The product may not work well when it is kept or operated beyond the temperature and

humidity specified.

If you want to turn off the BCM-600 while you’re operating, please use the power switch

where is the front of the product.

When you keep or operate the product in large temperature difference, product

malfunctioning can be happened because of humidity of battery pack. 

Please operate product after keep it room temperature for three hours.

Leakage, heating, ignition or rupture of the battery may cause fire or injury. 

Do not use batteries beyond those specified in this manual.

Do not short-circuit the battery and do not heat, disassemble, or dispose of it in a fire. 

Do not mix using the used battery with new batteries , different kind of batteries, other

company batteries  

Do not insert batteries with (+) and (-) ends reversed.

When the product is not used for a long time, separate the battery from the product.

To ensure the lifecycle and safety of battery, please  discharge the battery and then charge

perfectly again every six months. 

Also time-worn battery exceed 24 months should be replaced.

When the product has been moved, make sure to turn it off and arrange the accessories in

order. Injured cords or cables may cause fire or electric shock.

Precautions for Electric Safety

To prevent electric noise during use, the product should be installed apart from dynamo, X-

ray equipment, broadcast equipment or portable cables. An inaccurate result may occur when

these equipment are placed near the product. BCM-600 requires an independent power circuit

and a stable earth connection. Sharing power with other electric equipment may result in

inexact results.

BCM-600 is classified as follows :

Type 1 for electric shock, CF grade 

Please do not use this product near inflammable anesthetics and diluents.

Calssification is Class IIb

The noise level is “B” Class according to the IEC/EN 60601-1 (Safety of Electric Medical

Product), and the noise redemption is “B” Level according to the IEC/EN 60601-1-2

(Electromagnetic Compatibility Requirements)

Protection grade against water is “IPX 0”. 

Please check the following points before use.

Is the product supplied with the appropriate power? (100~240VAC)

Are all connecting parts (power cables or the product) properly connected?

Is the product completely grounded? (If not, noise may occur.)

Before turning the power on, is the accessory for measuring the target parameter well connected

to the product?

Summary of Contents for BCM-600

Page 1: ...82 33 434 9041 Fax 82 33 434 9043 Seoul Office R D 1406 Masters Tower 553 Dohwa dong Mapo gu Seoul Korea zip 121 748 Tel 82 2 714 2960 2 Fax 82 2 714 2963 Customer Service Dept Tel 82 2 714 2962 Fax 8...

Page 2: ...OPERATION MANUAL BCM 600 Electrocardiograph Ver 1 0 OCT 26 2010...

Page 3: ...M 600 10 8 Composition of BCM 600 11 9 How to set BCM 600 12 Precations for Settings 12 Checkpoints Before Use 12 Maintenance and Cleaning After Use 13 10 Components of BCM 600 14 Front Operation Pane...

Page 4: ...t that is not specified in the manual please be sure to notify our company or an agency authorized to distribute our products For various services and product information please feel free to contact o...

Page 5: ...nt all kinds of service related responsibilities of the product sellers for the product they sell The product was manufactured under strict quality controls and the inspection process of the company T...

Page 6: ...ipment Make sure to prevent dust especially metal debris from penetrating in Avoid damp locations and do not operate the equipment with wet hands Locations where the temperature fluctuation is rather...

Page 7: ...d 24 months should be replaced When the product has been moved make sure to turn it off and arrange the accessories in order Injured cords or cables may cause fire or electric shock Precautions for El...

Page 8: ...it prints in 4inch sized paper Also it can be easily used for an emergency or transport situation by built in battery and portable weight Features 12 channel ECG waves are configurated with 3channels...

Page 9: ...uch as battery LCD or LED and so on have a problem 1 Precautions for Settings Pay attention when setting BCM 600 for the following points Use BCM 600 under conditions of 10 45 of circumstantial temper...

Page 10: ...n 1 AC Power Button 2 Screen Shift Button 3 FREEZE Button 4 GAIN Button 5 PRINT START Button 6 SPEED Button 7 EXIT Button 8 MENU Button 9 LCD SCREEN 10 PAPER LOCK 11 PAPER Cover 12 AC Power LED 13 DC...

Page 11: ...nting and so on Refer the following page Table 1 Explanation of Screen Icons Time Display the present time HR Display the Heart Rate of a Patient Fig1 Main Screen Symbol Comments Symbol Description Ty...

Page 12: ...rmal Green Abnormal Red Normal Green Abnormal Red Flickers at 1 second interval Flickers at 1 second interval Flickers at 1 second interval Fixed Fixed How to Use BCM 600 1 How to Use Step 1 Turn on t...

Page 13: ...ft the menu button and press 1 EKG setting MENU MENU PRINT PATIENT SETUP EXIT MENU MENU PRINT PATIENT SETUP EXIT EKG LEAD SPEED GAIN FILTER MUSCLE BASELINE NOTCH Return Electrocardiograph Ver 1 0 20 M...

Page 14: ...s set NOTCH MONITOR DIAGNOSIS FREEZE MONITOR Patient s information WAVE display DIAGNOSIS Patient s information WAVE Diagnosis information FREEZE Patient s information WAVE output Changeable Range Des...

Page 15: ...TE TIME Initialize LANGUAGE Within 16 characters Year Month Day Hour Min OFF ON English Hospital name setting Date time setting Initialization Default Set referrence DEMO MODE setting Language Select...

Page 16: ...makes users to get more accurate judgments In order to change the print processing speed size and cannel setting press the Menu button to EKG menu where you can change the setting value If you use an...

Page 17: ...2 In order to change the print properties press the Menu button after moving to that properties by rolling the Menu button to right or left 3 To print select Print Menu And to move to no 1 select CANC...

Page 18: ...00 times 100 Charge discharge 50 Charge discharge 30 Charge discharge Power Battery Specifications During operation using AC power noise may occur in the waveform If this is the case ground the produc...

Page 19: ...B Electrode Attachment Check the status of the electrode attachment When you measure a patient s condition Make sure if there are mechanical hazards or not Check the cable connected to External and a...

Page 20: ...ble 3 Check if the product is grounded The power of the product is not turned off The waveform is not normal The power of the product turns off suddenly during use The power of the product is not turn...

Page 21: ...re Polarization voltage 350mV min Time constancy 3 2 seconds 0 20 Sensitivity 5 10 15 mm mV 5 Frequency response 0 05 100Hz 13dB Filters AC50 60 Hz 20dB or lower Muscle tremor 35 to 45 Hz 3db 6db oct...

Page 22: ...medical device Only the products that have Passed our extensive and thorough quality tests are offered to our customers Our policies on repairing exchange and refund complies with Regulation on Consu...

Reviews: