2.9 DEEP TISSUE HANDPIECE (OPTIONAL ACCESSORY)
NOTE:
The Deep Tissue Handpiece is reusable and equipped with a disposable non-sterile
protective shield for single patient use. The Handpiece is non-sterile and requires
cleaning before and after each patient treatment.
This Handpiece cannot be
sterilized in the autoclave.
For instructions on cleaning the Handpiece, refer to
section 8.
Always wipe the disposable shield with alcohol prior to use. The disposable shield is
for single-use only to avoid cross-contamination. Dispose of when treatment session
is completed.
Figure 2.15: Deep Tissue Handpiece
•
Remove Red Cap Dust Cover from the Deep Tissue Handpiece.
The Handpiece is now ready to use.
To remove the Handpiece, press and hold the buttons on the side of the Fiber Shaft and pull the
Handpiece away from the shaft.
•
Slide Handpiece over shaft until it clicks into place
(Figure 2.16).
Figure 2.16
•
Place protective shield over the adjustable spacer
(Figure 2.17).
Figure 2.17
•
Loosen the Lock Ring and set the Spacer at the desired spot
size Detent Location (Figure 2.18). Tighten the Lock Ring.
Figure 2.18
Handpiece
Spacer
Disposable
Shield
Red Cap Dust
Cover