8
5.1. Charging with the Biohit
e
LINE Charging
Stand and Carousel
The Biohit
e
LINE Charging Stands and
Carousels are bench top charging units.
The Charging Stand accommodates one
Biohit
e
LINE pipette and the Charging
Carousel four Biohit
e
LINE pipettes. The
e
LINE Charging Stands and Carousels
operate through the metal contacts placed
in the sides of the pipette and in the head
of the Charging units (Fig. 5.).
Before operating the AC-adaptor must be
connected to the socket at the rear of the
Charging Stand (Fig. 5) or at the side of
the Charging Carousel as well as to the
electrical outlet. Before connecting the
AC-adaptor to the electrical outlet, check
that its voltage settings corresponds to the
voltage settings of your electrical outlet.
Cat. No.
Product
730981
Biohit
e
LINE Charging Stand
730991
Biohit
e
LINE Charging Carousel
With universal AC-adaptor (Euro, USA/Jpn, UK and China plugs)
1. Check that the battery protector is removed and the battery cover is
properly closed.
2. Place the pipette into the Charging unit. Make sure that the metal
contacts at the rear of the pipette are properly
fi
tted to the tracks of the
metal contacts in the top of the Charging unit.
3. When charging, the battery symbol will be displayed with the + and –
symbols
fl
ashing. The charging time for completely discharged batteries
is 4 hours.
4. When the battery is full the battery symbol disappears from the display.
If display shows RESET and PRESS TIP, press the tip ejection button to
reset the pipette, the pipette will then be ready for use.
Note:
Never reset the pipette with the tip attached as the attached tip can
induce an error in the home position determination of the piston and tip
ejector collar!
Note:
When the pipette is not in use it is recommended to keep the pipette
in the Charging unit. This ensures that the battery is kept charged and the
pipette is ready for use.
Fig. 5.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com