13
▲
▼
7. Press or
to con
fi
rm selection and to display the speed out
8. Select the speed out with
(1= Min and 9 = Max)
9. Press or
to con
fi
rm selections
10. Position the tip to aspirate and press the START button
11. Position the tip to dispense. Press the START button
to dispense
the sample. The tip is emptied with blow-out and the piston returns
automatically back to the home position.
Step 10. Step 11.
Note:
When the START button is held down during dispensing the piston will
stop in the lowest position. When the start button is released the piston will
automatically return to the home position.
To mix:
1. Position the tip in to the solution, then press and hold the START button
.
Mixing is performed automatically as long as the START button
is
held down.
2. Position the tip to dispense. Press START button twice
Note:
The mixing is performed with about 70% of the total volume.
7.2. Reverse pipetting (rP)
A selected volume plus and excess is aspirated into the tip. After dispensing
the selected volume the excess volume remains in the tip and is discarded.
1. Press to display PIPET
2. Scroll
with until the symbol rP is displayed.
3. Con
fi
rm by pressing
4. Select the desired pipetting volume by using
Note:
When holding down
the volume display starts to scroll faster.
5. Press to con
fi
rm selection and to display the speed in
or
press and continue from Step 10.