LifePro Thermal Cycler Attachment 2
-
42
-
Attachment 2 Value and Meaning of Parameter Setting
No Parameter
Meaning
Scope of setting
value
Note
1 Temp(C)
Temperature
control point
(Celsius)
4.0~99.0
--:--
(press plus/minus to
set before :)
Temperature control time
is infinitely long, close
hot-lid at same time
2 Time
Temperature
control time
(mm:ss)
00
:
00~59
:
59
0.1 ~4.0
3 Ramp(C/S)
Rate of
temperature
rise/fall
(Celsius/second)
#.#
Maximum rate of
temperature rise/fall
4 +Temp
Temperature
increment in each
cycle (Celsius)
0.0~9.9
5 +Time
Time increment in
each cycle (m:ss)
0:00~9:59
Cycle1 Paragraph
number
1~5
×00
Cycle times in this
paragraph
1~99
From 00
First segment in
this paragraph
X (any segment
number already
existed)
6
to 00
Last segment in
this paragraph
≤
X+15
Cycle1
:
×00 From 00 to
00
7 Grad.
Gradient
temperature
(Celsius)
0 ~30
8 Date Date
(YY-MM-DD)
00
(
~99
)
-01
(
~12
)
-01
(
~31
)
Year-month-day
9 Time Time
(HH:MM:SS)
00:00:00~23:59:59
Hour: minute: second
10 Default
File
Default file name
in system
11 characters in
maximum length
11 Default
User
Default user name
in system
11 characters in
maximum length
Block
In this mode (it comes to
constant temperature
status once it reaches
target temperature), the
time that reagent
temperature at target
temperature is less than
time set (see Fig.1)
12 Control
Mode
Temperature
control mode
Tube
In this mode (it comes to
constant temperature
status after 3°C
overshoot once it
reaches target
temperature, the reagent
temperature approaches
target temperature quite
fast (see Fig.2)
13
Sample Volume
Sample size (ul)
10~200