12
Declaration of Conformity
Emissions
Manufacturer’s declaration on electromagnetic emissions
The Ultrasound Table is intended for use in the electromagnetic environment specified below. The customer or the user of the
Ultrasound Table should assure that it is used in such an environment
Emission test
Compliance
Electromagnetic environment
RF emissions
Group 1
The Ultrasound Table generates RF energy only for its internal func-
CISPR 11
tions. Therefore, its RF emission is very low and is not likely to cause
any interference in nearby electronic equipment
RF emissions
Class BB
The Ultrasound Table is suitable for use in all establishments
CISPR 11
including domestic establishments and those directly connected to the
public low-voltage power supply network supplying buildings used for
domestic purposes.
Immunity
Manufacturer’s declaration on electromagnetic immunity
The Ultrasound Table is intended for use in the electromagnetic environment specified below. The customer or the user of the
Ultrasound Table should assure that it is used in such an environment.
Immunity test
IEC 60601-1-2
IEC 60601-1-2
Test level
Compliance level
Electrostatic discharge
± 4 kV contact
Contact±4
(ESD)
± 8 kV air
kV Air ± 8 kV
IEC 61000-4-2
Electromagnetic environment – guidance
Floor should be wood, concrete or ceramic
tiles. If floor is covered with synthetic
material, the relative humidity should be at
least 30%