1
1. INTRODUCTION
Intended Use
The Econo Echocardiography Table is designed to provide a safe ergonomic environment for the
sonographer and patient. The design features are intended to relieve musculoskeletal stress on
the sonographer while scanning, and to provide the utmost patient comfort.
Indications For Use
The Econo Echocardiography Table is typically used
in radiology
departments, cardiovascular
facilities, diagnostic imaging centers and private practice clinics to achieve quality images for a
variety of procedures and patient applications.
General Cleaning and Maintenance
1. As required, cleanse all exterior surfaces and tabletop pads with a mild detergent solution
such as Parker Laboratories Protex Disinfectant or any one-step disinfectant that does not
contain bleach.
2. Keep wheel assemblies free of foreign materials and dirt accumulation.
3. Periodically inspect all welds.
4. Periodically check bolts on table, tighten if necessary.
5. Periodically inspect all strap holders. Any holder that feels loose should be removed and
reattached after adding a spot of Loctite
®
Threadlocker (Blue #242) to the bolt threads. This
should ensure that the bolt will not loosen again in the future.