background image

B

IODEX

Biodex Medical Systems, Inc.

ECONO 
ECHOCARDIOGRAPHY

 

TABLE

 

OPERATION MANUAL 

058-701 
058-706 

FN: 09-181 Rev C 5/15 

Summary of Contents for ECONO

Page 1: ...BIODEX Biodex Medical Systems Inc ECONO ECHOCARDIOGRAPHY TABLE OPERATION MANUAL 058 701 058 706 FN 09 181 Rev C 5 15...

Page 2: ...Econo Echocardiography Table This manual covers operation procedures for the following products 058 701 Table Econo Echocardiography 115 VAC 058 706 Table Econo Echocardiography 230 VAC...

Page 3: ...CHAPTER 2 ASSEMBLY AND TABLE OPERATION 3 Relocating the Table 3 Height Adjustment 3 Fowler Back Adjustment 3 Leveling The Table 3 Adjustable Sonographer Cutout Cushion 4 Cardiac Scanning Cutout 4 Inst...

Page 4: ...refully read these instructions prior to use Caution General Warning General Mandatory Action Dangerous Voltage On Power Off Power Earth ground Alternating Current Fuse USB Connector Cable Waste in El...

Page 5: ...r l quipement peut tre diminu e et les r sultats pourraient tre compromis Ne quittent Jamais le patient sans surveillance sur la table CAUTION Unauthorized modifications to this product are not permit...

Page 6: ...5th percentile male 20 65 years of age Product Certifications and Classifications The Econo Echocardiography Table has received the following certifications and falls within the following classificati...

Page 7: ...his manual This product should be used only as specified in the operation manual Biodex Imaging Tables are designed for use in a patient environment Tables Biodex imagerie sont con us pour une utilisa...

Page 8: ...e d connexion du produit d alimentation Ne placez pas le produit dans une position o le bouchon n est pas facilement accessible CAUTION This product is intended to remain in one location during operat...

Page 9: ...LUDING LOSS OF USE SALES PROFITS OR BUSINESS INTERRUPTION B This warranty does not apply if the product as determined by BIODEX MEDICAL SYSTEMS INC is defective due to abuse misuse modification or ser...

Page 10: ...ck of time 2 After taking the information we will advise on the action we will take 3 Sometimes service personnel must consult with engineering and it may take time to get back to you Be sure to let t...

Page 11: ...s for a variety of procedures and patient applications General Cleaning and Maintenance 1 As required cleanse all exterior surfaces and tabletop pads with a mild detergent solution such as Parker Labo...

Page 12: ...ty 2 Restraining Straps 3 Adjustable Sonographer Cutout Cushion 4 Fowler Back 5 Cardiac Scanning Cutout Cushion 6 Cardiac Scanning Cushion Release Lever 7 5 Locking Swivel Casters 8 Hand Controller No...

Page 13: ...iently adjusted with the hand held controller D ring holders are provided on the end of the straps at each side of the table frame to hold the hand held controller when it is not in use The hand held...

Page 14: ...hion two thirds of the way to vertical and then allow it to slide back down slightly until you feel it lock into position To raise the cushion to the fully raised position lift the cushion all the way...

Page 15: ...tube located on each side of the table below the Fowler back section The arm board pivot is adjustable at any point along its 130 arc To adjust press the yellow button on the arm board and move to the...

Page 16: ...of the cushion so the four pre drilled screw holes in the bracket align with the screw holes under the seat cushion 3 Using a 7 16 wrench install the four screws removed in step 1 through the side ra...

Page 17: ...st is removable and adjustable To install the headrest simply slide it into the receiving tube at the head end of the table until it clicks into place Figure 8 The optional headrest Figure 9 Headrest...

Page 18: ...secure the paper dispenser bracket in place 5 Repeat steps 1 4 for the opposite side 6 Place a role of paper on the paper dispenser rod and install the rod between the brackets Press in on either spri...

Page 19: ...inst the side frame so the two pre drilled screw holes in the bracket align with the screw holes on the tabletop frame The I V pole should face up and toward the head end of the table 3 Using a 7 16 w...

Page 20: ...ower body section 3 7 6 cm thick foam rubber pad covered with seamless durable Naugahyde Upholstery Color Graphite Motions Height Adjustable 23 to 39 58 4 to 99 cm Fowler Back 0 to 80 infinitely adjus...

Page 21: ...accessories transducers and cables sold by the manufacturer of this equipment as replacement parts for internal and external components may result in increased emissions or decreased immunity of the e...

Page 22: ...ble is suitable for use in all establishments CISPR 11 including domestic establishments and those directly connected to the public low voltage power supply network supplying buildings used for domest...

Page 23: ...range b Interference may occur in the vicinity of equipment marked with the following symbol Note 1 UT is the a c mains voltage prior to application of the test level Note 2 At 80 MHz and 800 MHz the...

Page 24: ...o 80 MHz 80 MHz to 800 MHz 800 MHz to 2 5 GHz d 1 2 P d 1 2 P d 2 3 P 0 01 0 12 0 12 0 23 0 1 0 38 0 38 0 73 1 1 2 1 2 2 3 10 3 8 3 8 7 3 100 12 12 23 For transmitters rated at a maximum output power...

Page 25: ...15 APPENDIX C PARTS AND ASSEMBLY ILLUSTRATIONS...

Page 26: ...16...

Page 27: ...17...

Page 28: ...18...

Page 29: ...19...

Page 30: ...20...

Page 31: ...21...

Page 32: ...22...

Page 33: ...BIODEX Biodex Medical Systems Inc 20 Ramsey Road Shirley New Y ork 11967 4704 Tel 800 224 6339 Int l 631 924 9000 Fax 631 924 9241 Email info biodex com www biodex com...