19
Biodex Medical Systems, Inc. © 2018
occurs, pedal/arm crank movement will gently cease and the screen will display an Overload!
message. A therapist is able to intervene and decide what action to take.
Figure 3.17.
Spasm control warning screen.
NOTE:
If pedal/arm crank rotation is completely resisted in the Passive Mode for a period of
time (about a minute), a “Motor over current protection” message may be displayed. To
protect the motor mechanism, the device will suspend operations. When this event occurs,
turn the device off with the ON / OFF switch at the base and turn the power back on.
Emergency Stop
The medBike is equipped with an <EMERGENCY STOP>. If, for any reason, a user wants to
quickly pause or discontinue an exercise session — be it in Active or Passive mode — it can be
accomplished by pressing <STOP>. To continue exercising, <STOP> must be pressed a second
time. There is an option to resume the session or finish the session and starting a new one.
Figure 3.18.
Location of Emergency Stop button.