
25
Biodex Medical Systems, Inc. © 2018
Figure 4.9.
Active Resistance Mode (Program A) Results.
Press <Home> to return to the Home screen.
Figure 4.10.
Home screen.
Continuous Watt Mode (Program B)
In Active Program B mode, the level of Watts generated by the user’s pedal/arm cranking is
fixed instead of the resistance level as with the Manual and Program A modes. Because Watts
are a function of rpms and resistance level, the two factors will adjust depending on the user’s
performance. Therefore, if the user is pedal/arm cranking quickly and generating a high
number of rpms, the resistance level will drop. If the user is producing a small number of rpms,
the resistance level will raise proportionately to maintain the designated Watt output.
Summary of Contents for 950-192
Page 1: ...medBike OPERATION MANUAL 950 192 950 194 950 193 FN 16 181 Rev B 1 17...
Page 20: ...20 Biodex Medical Systems Inc 2018 Figure 3 19 Emergency Stop warning screen...
Page 28: ...28 Biodex Medical Systems Inc 2018 Figure 4 15 Home screen...
Page 36: ...36 Biodex Medical Systems Inc 2018 Figure 5 15 Home screen...
Page 38: ...38 Biodex Medical Systems Inc 2018 Calf Support Assembly...