biochrom GeneQuant 1300 Quick Reference Manual Download Page 8

 

 

 

GeneQuant

TM

 1300 Quick Reference Guide  Page 8 

Version 2.0 

 

 

Parameter 

Folder 

Sub-Folder 

Manual 

page 

Description and options 

Units 

App

lications 

 

 

Life Science 

Concentration 

Kinetics – Parameters 2 

Standard Curve 

Protein – BCA 

Protein – Bradford 

Protein – Lowry 

Protein – Biuret 

45 

50 

53 

30 

33 

36 

39 

Enter the units using the alphanumeric keys 

or press 

::;

  and select pre-defined units 

using the left and right arrows (options: 

(µg/ml, µg/µl, pmol/µl, mg/dl, mmol/l, µmol/l, 

g/l, mg/l, µg/l, U/l, %, ppm,  ppb, conc or 

none) 

Units 

Life Science  OD600 

25 

Select the units of measurement: OD or 

cells/ml 

Volume 

App

lications 

 

Life science 

Absorbance Ration – 

Parameters 

DNA 

RNA 

Oligo 

Cy Dye 

Protein – Protein UV 

58 

 

13 

15 

17 

23 

28 

Enter the volume of the sample. Range:  

0.01 to 9999 

Volume 

Life science  Cy Dye 

23 

Enter the volume of the probe being used in 

µl 

Wavelength 

Applications  Concentration 

45 

Enter the wavelength at which you want to 

do the colorimetric assay 

Wavelength 

Applications  Kinetics – Parameters 1 

50 

Enter the wavelength at which you want to 

measure absorbance over a period of time 

Wavelength 

Applications  Single wavelength 

43 

Enter the wavelength at which you want to 

measure absorbance or % transmission 

Wavelength 

Applications  Standard curve 

53 

Select the wavelength at which you want to 

construct the calibration curve 

Wavelength 

Life science  Cy Dye 

23 

Enter the wavelength of the dye absorbtion 

peak 

Wavelength 

Life Science  OD600 

25 

Select the wavelength. Default value is 600 

nm  

Wavelength 

Life Science  Protein – BCA  

30 

Set at 562 nm  

Wavelength 

Life Science  Protein – Biuret 

39 

Set at 546 nm 

Wavelength 

Life Science  Protein – Bradford  

33 

Set at 595 nm  

Wavelength 

Life Science  Protein – Lowry  

36 

Set at 750 nm  

Wavelength 1 

Applications  Absorbance Ratio – 

Wavelengths 

58 

Enter the first wavelength which you want to 

use to measure the absorbance ratio 

Wavelength 2 

Applications  Absorbance Ratio – 

Wavelengths 

58 

Enter the second wavelength which you 

want to use to measure the absorbance 

ratio 

Wavelength 3 

Applications  Absorbance Ratio – 

Wavelengths 

58 

Enter the wavelength from which the 

background correction will be obtained. This 

parameter is only available if the 

background parameter has been set to On 

Wavelengths 

Applications  Multi Wavelength 

57 

Select the number of wavelengths at which 

you want to measure absorbance. Range 2-

X axis limits 

Applications  Wavescan, options 6 - 

Graph Scale 

49 

Set to On to define the start and finish 

points of the x axis, or off to retain default 

values 

X1 

Applications  Wavescan, options 6 - 

Graph Scale 

49 

Enter the minimum value for the x axis 

X2 

Applications  Wavescan, options 6 - 

Graph Scale 

49 

Enter the maximum value for the x axis 

Summary of Contents for GeneQuant 1300

Page 1: ...ntTM 1300 QUICK REFERENCE GUIDE Biochrom US Telephone 1 508 893 8999 84 October Hill Road Toll Free 1 800 272 2775 Holliston MA Fax 1 508 429 5732 01746 1388 support hbiosci com USA www biochromspectr...

Page 2: ...parameters and to write text descriptions where appropriate or required Use repeated key presses to cycle through lower case number and upper case Leave for 1 second before entering next character Use...

Page 3: ...s the symbol either type in a value or press the options key and choose a parameter from the next screen OR If the box contains arrow symbols use the left and right arrow keys to select the required p...

Page 4: ...vlengths Background Life Science Nucleic acids DNA Nucleic acids RNA Nucleic acids Oligo Cy Dye Protein Protein UV 13 15 17 23 28 Select whether the background correction at 310 nm is used or not Opti...

Page 5: ...meters 2 Standard Curve Protein BCA Protein Bradford Protein Lowry Protein Biuret 45 50 53 30 33 36 39 Determines the number of decimal places in the results 0 2 Results have a maximum of 5 figures Dr...

Page 6: ...f the two adjacent minima that a peak must be if it is to be detected Minimum peak width Applications Wavescan options 4 peak detection 48 This selects the minimum width in nm a peak must be to be det...

Page 7: ...Options OFF 1 2 or 3 This parameter is only available if the calibration mode is set to Standards Sort peaks by Applications Wavescan options 4 peak detection 48 Select how peaks are sorted by wavele...

Page 8: ...lect the wavelength at which you want to construct the calibration curve Wavelength Life science Cy Dye 23 Enter the wavelength of the dye absorbtion peak Wavelength Life Science OD600 25 Select the w...

Page 9: ...options 6 Graph Scale 49 Enter the minimum value for the y axis Y2 Applications Wavescan options 6 Graph Scale 49 Enter the maximum value for the y axis Year Utiliti Date and T 62 Enter the year Zoom...

Reviews: