11
6.5.3 Select CHECK.
6.5.4 Take a check strip out of the container.
IMPORTANT :
Do not touch the surface
of the check strip. A check strip contaminated
by sebum may cause incorrect results.
6.5.5 Put a Check Strip in the strip holder.
6.5.6 Press Quick button
to initiate
CHECK STRIP analysis.
6.5.7 Upon the completion of analysis, the
strip holder will eject automatically. The
machine will print out the analysis
report automatically.
Pass
means the
machine passed the test;
Fail
means
the machine failed the test.
<Strip Type>
S-10L S-11 S-11VET CHECK
Date :2018/04/19 17:38
SEQ# : CHECK
ID# :Check Strip Test
MEMO
- - - - - - - - - - - - - - - - - -
CHECK : Pass