28
Fig. 5-2
Step 5, Reset the Fuse Seat
Replace the fuse seat in the power socket.
Step 6, Switch On the Power
Plug the socket and switch on the power.
(2) Replace Lamps
High temperature
!
Wait 20 minutes before open the lamp chamber after power off to
avoid scald!
Step 1, Tools Preparation
Prepare a 6×150mm Cross Blade screwdriver and a pair of glove.
Step 2, Power Off
Switch off the power supply and unplug the socket.
Step 3, Open the Cover
Unscrew the 4 screws indicated in Fig.5-3(Each side with 2 screws)and remove the cover.
Fig. 5-3
Step 4, Open the Cover of the Light Chamber
Unscrew the 2 screws on the light chamber cover and remove it(Fig.5-4).