Recommended Test Schedule
|
105
Recommended Test Schedule
The schedule shown in
Table 1
defines the factory-recommended intervals for
performance testing for a microplate reader used two to five days a week.
Note:
The risk factors associated with your tests may require
that the Performance and Operational Qualification procedures
be performed more or less frequently than shown below.
Table 1
Recommended Test Schedule
Installation
Qualification
Performance
Qualification:
Monthly
Performance
Qualification:
Semiannually
Operational
Qualification:
Initially and
Annually
Absorbance Plate Test,
p. 110
Liquid Test 2, p.121
*
*
Liquid Test 3, p. 125
Optional for 340 nm
*Run Liquid Test 2
ONLY
if you do not have an Absorbance Test Plate.
If you run Liquid Test 2, you do
NOT
have to also run Liquid Test 1.
ELx800 Operator's Manual
Summary of Contents for ELx800
Page 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Page 2: ......
Page 4: ......
Page 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Page 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Page 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Page 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Page 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Page 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Page 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Page 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Page 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Page 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Page 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...