Liquid Testing
|
121
Liquid Test 2
The recommended method of testing the instrument performance is to use
the Absorbance Test Plate to confirm alignment, repeatability, and
accuracy, which will also confirm linearity.
If a Test Plate is not available,
Liquid Test 2
can be used for these tests.
Required Materials:
•
New 96-well, flat-bottom microplates (Corning Costar® #3590 is
recommended)
•
Ten test tubes, numbered consecutively, stored in a rack
•
Calibrated hand pipette (Class A volumetric pipette
recommended)
•
Stock solution A or B (these are the same solutions as for Liquid
Test 1)
Preparation of Dilutions:
1
Set up a rack containing 10 tubes, numbered consecutively.
2
If you have not already done so, prepare concentrated stock Test
Solution A or B.
Refer to
Table 3
when executing steps 3 and 4.
Table 3
Test Tube Dilutions
Tube Number
1
2
3
4
5
6
7
8
9
10
Volume of Concentrated
Solution (ml)
20
18
16
14
12
10 8 6
4
2
Volume of 0.05% Tween
Solution (ml)
0
2
4
6
8
10 12 14
16
18
Absorbance expected if
concentrated solution is
2.0 at 200
µ
l
2.0
1.8
1.6
1.4
1.2
1.0 0.8 0.6
0.4
0.2
3
Create a percentage dilution series, beginning with 100% of the
concentrated stock solution in tube 1, 90% of the concentrated
solution in tube 2, 80% in tube 3, and so on to 10% in tube 10.
4
Dilute using amounts of the remaining 0.05% solution of
deionized water and Tween 20, as shown in
Table 3
.
ELx800 Operator's Manual
Summary of Contents for ELx800
Page 1: ...ELx800 Absorbance Microplate Reader Operator s Manual...
Page 2: ......
Page 4: ......
Page 40: ...12 Chapter 1 Introduction Bio Tek Instruments Inc...
Page 66: ...38 Chapter 2 Installation Bio Tek Instruments Inc...
Page 130: ...102 Chapter 3 Operation Bio Tek Instruments Inc...
Page 166: ...138 Chapter 5 Maintenance and Decontamination Bio Tek Instruments Inc...
Page 190: ...162 Appendix B Using 384 Well Geometry Bio Tek Instruments Inc...
Page 193: ...165 Figure 32 Curve Fit Report ELx800 Operator s Manual...
Page 195: ...167 Figure 34 Column Report without samples ELx800 Operator s Manual...
Page 196: ...168 Appendix C Report Format Figure 35 Panel Report Bio Tek Instruments Inc...
Page 197: ...169 Figure 36 Assay Detail Report Sheet 1 of 2 ELx800 Operator s Manual...
Page 200: ...172 Appendix D Comparison of the ELx800 and the EL800 Bio Tek Instruments Inc 172...
Page 214: ...186 Appendix E Instructions for Programming a New Assay Bio Tek Instruments Inc...