BIO RAD Ultramark User Manual Download Page 9

2.  Equipment Set-up and Installation 

2.1 

Unpacking the Ultramark Microplate Reader 

Before opening the box, check for any outer damage. Document any damages, if present. 

Place the box upright with the TOP label facing up and open the box. Remove  the instrument from 
the box in this position only. 

Warning!

 Grasp the instrument firmly under the right and left sides before attempting to lift or carry 

it. 

Check the instrument for damaged or broken parts. Document any damages in detail. 

Each Ultramark instrument includes the following accessories. Verify that they are present: 

• 

1 power cord. 

• 

1 plastic container with filter holder and 10 filters: 260 nm, 280 nm, 340 nm, 405 nm, 415 nm, 
490 nm, 550 nm, 630 nm, 655 nm, and 750 nm. 

• 

1 user manual. 

Verify that the serial number on the back of the instrument is the same as that on the outside of the 
box.  

Remove the side cover on the right side of the unit. Inside there are two docking brackets that need to 
be removed by unscrewing 3 screws as shown on the unpacking sheet attached to the unit. After 
removing the two docking brackets, install the Filter Magazine as described in the unpacking 
instructions. 

2.2 

SCSI II Cable Connection 

The SCSI II port is located on the back of the instrument. There are two connectors, which allows the 
user to connect more than one SCSI instrument in a chain. If the Ultramark is the last instrument in 
the chain, make sure that the termination switch (next to the SCSI connectors) is in the “ON” position. 
If the Ultramark is an intermediate unit in the chain, make sure that the termination switch is in the 
“OFF” position. If the Ultramark is the only unit connected to the SCSI cable, make sure that the 
switch is in the “ON” position. 

There is a small dial next to the SCSI connectors on the back of the unit. This dial selects the SCSI ID 
number for the unit at boot-up. Be sure to select an ID number that is not being used by any other 
SCSI device in the chain. 

2.3 Power 

Connection 

Before switching the power on for the first time, allow the Ultramark to warm to room temperature 
for 

6 hours

 after unpacking to prevent any condensation problems. 

Plug the included power cord into the main power connector port. Make sure that the instrument is 
connected to a grounded conductor of the main power outlet. 

Summary of Contents for Ultramark

Page 1: ...410 0164 Rev B Ultramark Microplate Reader User Manual...

Page 2: ...S LSG TechServ Intl Bio Rad com International Notice No part of this publication may be reproduced or transmitted in any form or by any means electronic or mechanical including photocopy recording or...

Page 3: ...particular purpose or otherwise expressed or implied in fact or by law and state Bio Rad s entire and exclusive remedy for any claims or damages in connection with the furnishing of goods or parts the...

Page 4: ......

Page 5: ...Connection 3 2 3 Power Connection 3 2 4 Location of the Instrument 4 3 Using the Ultramark 5 3 1 Loading a Plate 5 3 2 Reading a Plate 5 3 3 Incubator Controls 5 4 Changing Filters in the Ultramark 7...

Page 6: ......

Page 7: ...Shaking Yes Incubation Up to 42 C programmable in 0 2 C increments Power 100 240 V 60 50 Hz 22 Amps Dimensions 20 cm H x 43 cm W x 51 cm D Compatible Plates Any microplate or suitable medium with dim...

Page 8: ......

Page 9: ...removing the two docking brackets install the Filter Magazine as described in the unpacking instructions 2 2 SCSI II Cable Connection The SCSI II port is located on the back of the instrument There ar...

Page 10: ...sitivity Place the instrument on a flat surface that is free of vibration Do not place the instrument in direct sunlight Make sure that the instrument is being operated in a dust free environment Warn...

Page 11: ...ing a Plate Plate readings are controlled entirely by Microplate Manager See Chapter 5 of the Microplate Manager User Guide for instructions To stop a plate reading that is in progress press the Stop...

Page 12: ......

Page 13: ...screw Screw 2 at the bottom of the Filter Magazine Pull the Filter Magazine forward off the pins and lift it out of the Filter Base Press any button on the front panel and the filter arm will extend R...

Page 14: ...into the base so that the pins in the base fit in the holes in the magazine Insert Screw 2 into the center hole in the Filter Magazine and tighten securely with a coin or screwdriver The Filter Magazi...

Page 15: ...the Filter Magazine from the Ultramark see instructions in previous section 3 Note the number of the filter that is in the filter arm i e the Magazine clip that does not have a filter 4 To enter the...

Page 16: ......

Page 17: ...11 6 Recovering a Dropped Filter To recover a filter that has dropped inside the instrument perform the following steps 1 Turn power off 2 Loosen thumbscrews on side panel and remove panel...

Page 18: ...lter magazine 5 Reach inside instrument locate dropped filter and remove 6 Reinsert filter into clip in magazine replace the magazine and side panel and turn the instrument back on 7 Specify in Microp...

Page 19: ...ment Without the Disinfection Certificate Bio Rad Laboratories will not accept any returned instruments Bio Rad Laboratories recommends using a solution of 5 sodium hypochlorite solution in water Auth...

Page 20: ......

Page 21: ...aged Contact Bio Rad X Axis Dirty guide rail or something blocking the rail Confirm that there is nothing blocking the guide rail then clean the rail and put grease on it Motor drive damaged Contact B...

Page 22: ...s 3 Xenon Flash Lamp damaged Contact Bio Rad Measuring electronics damaged Contact Bio Rad Optical Block is out of the starting point Contact Bio Rad Lenses and Mirrors may be dirty Contact Bio Rad Ga...

Page 23: ...Rad SCSI bus Data communication problem between the Ultramark and your computer The Ultramark is not the source of the problem Check your computer Bus Rst SCSI The termination switch may be off Turn t...

Page 24: ......

Page 25: ...ansparent Microplate 25 170 9520 SCSI II Card and Cable for PC 170 9521 MPM 5 0 PC Software Filter Table 170 9504 260 nm Filter 170 9505 280 nm Filter 170 9506 340 nm Filter 170 9507 405 nm Filter 170...

Page 26: ......

Page 27: ...from cut on to UV Edge slope Per Omega published data for EF series Each filter whose construction incorporates absorbing glass shall be oriented with the absorbing glass facing the rim of the alumin...

Page 28: ......

Page 29: ...d its inventory have 1 never been in contact with any dangerous biological material or 2 been disinfected according to the instructions contained in the operating manual for this instrument Name pleas...

Page 30: ......

Page 31: ...25...

Reviews: