3.  Using the Ultramark 

The Ultramark is operated through the Protocol dialog boxes in Microplate Manager. See the 

Microplate Manager User Guide 

for instructions on specifying the filter wavelengths (section 2.7) 

and operating the Ultramark through the Protocol dialogs (Chapter 5). 

3.1 

Loading a Plate 

To load a plate into the Ultramark, press the 

Open/Close button 

on the front of the instrument or 

click on the Reader Door: Open button in the Protocol dialog box in Microplate Manager. The front 
door of the reader will open and the plate loader will extend. 

Place your microplate in the plate loader, pressing lightly to snap the plate into place. Press the 
Open/Close button again to close the reader door, or click on the Reader Door: Close button in the 
Protocol dialog in Microplate Manager. 

 

The Ultramark will accept any plate with the dimensions 138 mm (length) x 98 mm (width). 
Plate height cannot exceed 15 mm. 

3.2 

Reading a Plate 

Plate readings are controlled entirely by Microplate Manager. See Chapter 5 of the Microplate 
Manager User Guide for instructions. 

To stop a plate reading that is in progress, press the 

Stop button

 on the front panel of the Ultramark 

or click on the Stop button in the Microplate Manager Protocol dialog box. 

3.3 Incubator 

Controls 

The incubator in the Ultramark is controlled from within Microplate Manager. You can also start the 
incubator by pressing the 

Incubator button

 on the front panel of the Ultramark. 

Pressing the Incubator button once will begin warming up the incubator (message: “Ready. Temp = 
__”). Pressing it again will begin cooling the incubator down to ambient room temperature (message: 
“Cooling”). Pressing it again will turn the incubator off (message: “Ready. Incu. Off”). 

Incubator Set Point 

The incubator set point is set from within Microplate Manager. The last set point selected in 
Microplate Manager is stored in memory inside the Ultramark, so if you start the Ultramark and start 
the incubator without opening Microplate Manager, the incubator will begin warming up to the last 
specified set point. 

 

Summary of Contents for Ultramark

Page 1: ...410 0164 Rev B Ultramark Microplate Reader User Manual...

Page 2: ...S LSG TechServ Intl Bio Rad com International Notice No part of this publication may be reproduced or transmitted in any form or by any means electronic or mechanical including photocopy recording or...

Page 3: ...particular purpose or otherwise expressed or implied in fact or by law and state Bio Rad s entire and exclusive remedy for any claims or damages in connection with the furnishing of goods or parts the...

Page 4: ......

Page 5: ...Connection 3 2 3 Power Connection 3 2 4 Location of the Instrument 4 3 Using the Ultramark 5 3 1 Loading a Plate 5 3 2 Reading a Plate 5 3 3 Incubator Controls 5 4 Changing Filters in the Ultramark 7...

Page 6: ......

Page 7: ...Shaking Yes Incubation Up to 42 C programmable in 0 2 C increments Power 100 240 V 60 50 Hz 22 Amps Dimensions 20 cm H x 43 cm W x 51 cm D Compatible Plates Any microplate or suitable medium with dim...

Page 8: ......

Page 9: ...removing the two docking brackets install the Filter Magazine as described in the unpacking instructions 2 2 SCSI II Cable Connection The SCSI II port is located on the back of the instrument There ar...

Page 10: ...sitivity Place the instrument on a flat surface that is free of vibration Do not place the instrument in direct sunlight Make sure that the instrument is being operated in a dust free environment Warn...

Page 11: ...ing a Plate Plate readings are controlled entirely by Microplate Manager See Chapter 5 of the Microplate Manager User Guide for instructions To stop a plate reading that is in progress press the Stop...

Page 12: ......

Page 13: ...screw Screw 2 at the bottom of the Filter Magazine Pull the Filter Magazine forward off the pins and lift it out of the Filter Base Press any button on the front panel and the filter arm will extend R...

Page 14: ...into the base so that the pins in the base fit in the holes in the magazine Insert Screw 2 into the center hole in the Filter Magazine and tighten securely with a coin or screwdriver The Filter Magazi...

Page 15: ...the Filter Magazine from the Ultramark see instructions in previous section 3 Note the number of the filter that is in the filter arm i e the Magazine clip that does not have a filter 4 To enter the...

Page 16: ......

Page 17: ...11 6 Recovering a Dropped Filter To recover a filter that has dropped inside the instrument perform the following steps 1 Turn power off 2 Loosen thumbscrews on side panel and remove panel...

Page 18: ...lter magazine 5 Reach inside instrument locate dropped filter and remove 6 Reinsert filter into clip in magazine replace the magazine and side panel and turn the instrument back on 7 Specify in Microp...

Page 19: ...ment Without the Disinfection Certificate Bio Rad Laboratories will not accept any returned instruments Bio Rad Laboratories recommends using a solution of 5 sodium hypochlorite solution in water Auth...

Page 20: ......

Page 21: ...aged Contact Bio Rad X Axis Dirty guide rail or something blocking the rail Confirm that there is nothing blocking the guide rail then clean the rail and put grease on it Motor drive damaged Contact B...

Page 22: ...s 3 Xenon Flash Lamp damaged Contact Bio Rad Measuring electronics damaged Contact Bio Rad Optical Block is out of the starting point Contact Bio Rad Lenses and Mirrors may be dirty Contact Bio Rad Ga...

Page 23: ...Rad SCSI bus Data communication problem between the Ultramark and your computer The Ultramark is not the source of the problem Check your computer Bus Rst SCSI The termination switch may be off Turn t...

Page 24: ......

Page 25: ...ansparent Microplate 25 170 9520 SCSI II Card and Cable for PC 170 9521 MPM 5 0 PC Software Filter Table 170 9504 260 nm Filter 170 9505 280 nm Filter 170 9506 340 nm Filter 170 9507 405 nm Filter 170...

Page 26: ......

Page 27: ...from cut on to UV Edge slope Per Omega published data for EF series Each filter whose construction incorporates absorbing glass shall be oriented with the absorbing glass facing the rim of the alumin...

Page 28: ......

Page 29: ...d its inventory have 1 never been in contact with any dangerous biological material or 2 been disinfected according to the instructions contained in the operating manual for this instrument Name pleas...

Page 30: ......

Page 31: ...25...

Reviews: