4 Press the first key left on the thermoregulator. The display shows the set working temperature (default -20°C).
Without releasing the first left button, press the arrow keys depending on the temperature you want to set.
5 Release all the keys. After a few minutes the actual working temperature will be the same as that set.
6 Replace the door in the seat and tighten all the screws.
7 Place the machine in its original position.
Installation
Position the instrument on a level and stable working bench and connect it to the power supply socket (230V~
50/60Hz) using the provided cable.
D
o not use any extension or adapter and do not modify the provided cable.
Important:
Make sure that the back grid has at least a 15 cm free space in order to allow the aeration of the
cooling system and the optimal system performance. In case of installing multiple units, never place the instruments
with rear aeration grids against each other. If it is not possible, leave a space of at least 60 cm between a rear grid
and the other.