10) CLEANING AND MAINTENANCE
Cleaning
A proper standard of hygiene and the routine maintenance of the laboratory are necessary to keep the instrument in
good working conditions. The operator is responsible for the proper cleaning of the hood.
1) Use only products suitable for steel surfaces. Do not use brushes with hard bristles nor abrasive sponges or
detergents in order to avoid ruining the surface.
2) Wear anti-cut gloves over or under the common gloves to avoid possible accidents.
3) The table is made of AISI 304 stainless steel, resistant against a lot of solvents and acids except for muriatic acid
that might cause a rust effect.
4) In case of accidental pouring of dangerous or corrosive material, remove every trace of substance immediately.
The prolonged contact of such substances with steel might stain it and, in extreme cases, corrode it.
5) The control panel keys are sealed by a membrane which prevent liquids or dirt from entering. It is recommended,
however, to clean periodically the panel surface with a soft and damp cloth and avoid the contact with
hydrocarbons and chemical substances (xylol, toluol, ecc.) which might damage it.
Maintenance
Daily
-
Wash the basins and the working surface using the extractable shower and suitable products.
-
Control the outlet of the aspirated basin placed under the perforated working surface, in order to avoid accidental
leaks of liquids in the aspiration chambers.
-
In case of use of the garbage disposal unit, observe the instructions in paragraph 9) Garbage disposal unit on
page 18.
Half-yearly
-
Control that the filter isn’t obstruct by formalin salts or dust and that isn’t consumed. Effect the replacement if
necessary. The table is, however, provided with visual and acoustic alarm in case of filters saturation.
Annually
Effect the following main operations:
-
Verify the wholeness of stainless steel structure (working surface and sinks), power supply cable and aspiration
tube.
-
Replace the filter apart form the use of the table.
-
Effect the electric controls inspecting control panel and fan.
-
Cleaning and general working.