10) CLEANING AND MAINTENANCE

 

 

Cleaning 

 
A proper standard of hygiene and the routine maintenance of the laboratory are necessary to keep the instrument in 
good working conditions. The operator is responsible for the proper cleaning of the hood. 
 
 
1)  Use  only  products  suitable  for  steel  surfaces.  Do  not  use  brushes  with  hard  bristles  nor  abrasive  sponges  or 

detergents in order to avoid ruining the surface. 

2)  Wear anti-cut gloves over or under the common gloves to avoid possible accidents. 
3)  The table is made of AISI 304 stainless steel, resistant against a lot of solvents and acids except for muriatic acid 

that might cause a rust effect. 

4)  In case of accidental pouring of dangerous or corrosive material, remove every trace of substance immediately. 

The prolonged contact of such substances with steel might stain it and, in extreme cases, corrode it. 

5)  The control panel keys are sealed by a membrane which prevent liquids or dirt from entering. It is recommended, 

however,  to  clean  periodically  the  panel  surface  with  a  soft  and  damp  cloth  and  avoid  the  contact  with 
hydrocarbons and chemical substances (xylol, toluol, ecc.) which might damage it.   

 
Maintenance 

 
Daily 
 

-

 

Wash the basins and the working surface using the extractable shower and suitable products. 

-

 

Control the outlet of the aspirated basin placed under the perforated working surface, in order to avoid accidental 
leaks of liquids in the aspiration chambers. 

-

 

In case of use of the garbage disposal unit, observe the instructions in paragraph 9) Garbage disposal unit on 
page 18. 

 
Half-yearly 
 

-

 

Control that the filter isn’t obstruct by formalin salts or dust and that isn’t consumed. Effect the replacement if 
necessary. The table is, however, provided with visual and acoustic alarm in case of filters saturation. 

 
Annually 
 
Effect the following main operations: 
 

-

 

Verify the wholeness of stainless steel structure (working surface and sinks), power supply cable and aspiration 
tube. 

-

 

Replace the filter apart form the use of the table. 

-

 

Effect the electric controls inspecting control panel and fan. 

-

 

Cleaning and general working. 

 
 
 
 
 
 
 
 

Summary of Contents for 50-270-701

Page 1: ...o S p A via San Faustino 58 I 20134 Milano Tel 39 02 21 27 13 1 Fax Acquisti Export 39 02 21 54 155 Fax Assistenza Contabilit 39 02 26 41 74 48 Fax Vendite 39 02 21 53 000 Revision 160613 of 13 06 16...

Page 2: ...y if it is correctly connected to an efficient earth circuit as indicated by current electrical safety regulations It is necessary to check this fundamental safety prerequisite and if in doubt ask to...

Page 3: ...f 13 06 16 Page 3 of 21 11 Graphic symbols indicated on the label positioned near the instrument s power supply socket Symbol for CATALOGUE NUMBER Symbol for SERIAL NUMBER Symbol for ALTERNATING CURRE...

Page 4: ...6 16 Page 4 of 21 INDEX 1 Technical features Page 5 2 Installation Page 7 3 Water and electric connections Page 7 4 Aspiration system Page 9 5 Replacement of filter Page 10 6 Control panel Page 12 7 F...

Page 5: ...ided with a Schuko universal power supply socket for the possible connection of external instruments Structural features Structure base collecting basin sink perforated planes headrest servant table m...

Page 6: ...39 02 26 41 74 48 Fax Vendite 39 02 21 53 000 Revision 160613 of 13 06 16 Page 6 of 21 Dimensional features Overall dimensions in mm Working plane usable dimensions in mm Sink dimensions in mm Weight...

Page 7: ...ischarged air 3 WATER AND ELECTRIC CONNECTIONS Water inlet The tap of the sink mixer can supply hot and cold water while the wash of the collecting basin can be connected only to cold water The mixer...

Page 8: ...ite 39 02 21 53 000 Revision 160613 of 13 06 16 Page 8 of 21 Plumbing diagram Electrical connections Before connecting the instrument to the electric system using the provided cable verify the adequac...

Page 9: ...ection to the hospital s centralized canalization system for vapours discharge outside The electrofan allows vapours aspiration from the working surface precisely from the table s perimeter and the ce...

Page 10: ...he socket 2 Wear individual protection devices in compliance with the regulation in force 3 Lift the table up to end run remove voltage by pressing the red emergency switch unscrew the knobs remove th...

Page 11: ...S p A via San Faustino 58 I 20134 Milano Tel 39 02 21 27 13 1 Fax Acquisti Export 39 02 21 54 155 Fax Assistenza Contabilit 39 02 26 41 74 48 Fax Vendite 39 02 21 53 000 Revision 160613 of 13 06 16 Pa...

Page 12: ...OL PANEL Control panel with polycarbonate soft touch keyboard and digital display for the working parameters visualization and modification Main functions General ignition switch Emergency switch Star...

Page 13: ...21 53 000 Revision 160613 of 13 06 16 Page 13 of 21 Description of keys and switches Display Visual and acoustic alarm SET Parameters setup change Key MUTE Alarms silencing key Values change keys Inc...

Page 14: ...Acquisti Export 39 02 21 54 155 Fax Assistenza Contabilit 39 02 26 41 74 48 Fax Vendite 39 02 21 53 000 Revision 160613 of 13 06 16 Page 14 of 21 Example of parameters change 3 Function 4 Confirm the...

Page 15: ...fan s self switching off but it can be zeroed by the operator by pressing at the same time the increase decrease keys placed over MENU key The count happens only if the fan is in operation Filter life...

Page 16: ...dite 39 02 21 53 000 Revision 160613 of 13 06 16 Page 16 of 21 How ELAPSED TIME of the function LIVE FILTER is zeroed After changing the filter zero ELAPSED TIME by pressing at the same time the incre...

Page 17: ...te are saved automatically on internal EPROM every hour and for every exit from the screen of a parameter by pressing the MENU key Alarms function When a determinate function reaches the maximum value...

Page 18: ...l for the starting Important 1 The garbage disposal unit must be absolutely used running abundantly water into the sink 2 In order to start it press the provided pedal 3 Do not insert metal or plastic...

Page 19: ...might stain it and in extreme cases corrode it 5 The control panel keys are sealed by a membrane which prevent liquids or dirt from entering It is recommended however to clean periodically the panel...

Page 20: ...microphone and pedal control 50 500 056 3 Stainless steel sink filter 50 500 062 4 Liquids aspirator 50 700 052 These accessories can be installed only at the time of production not afterwards 12 MAIN...

Page 21: ...ice das Medizinprodukt f r die In vitro_Diagnostik le dispositif m dical de diagnostic in vitro TAVOLO AUTOPTICO POSTMORTEM TABLE TV410 della classe of class der Klasse de la classe Altro Other Sonsti...

Reviews: