8) REPLACING THE FUSES
If the green switch is on but the table doesn’t work, check, and if necessary replace, the delayed fuse placed inside
the control panel: 3,15 A 5x20 mm – T3.15AH250V.
9) GARBAGE DISPOSAL UNIT (OPTIONAL ACCESSORY)
The garbage disposal unit is used to avoid possible obstructions in the draining pipes. It is provided with stainless
steel working chamber and rotating parts and a pedal control for the starting.
Important:
1) The garbage disposal unit must be absolutely used running abundantly water into the sink.
2) In order to start it, press the provided pedal.
3) Do not insert metal or plastic objects.
4) Do not insert the hands in the working chamber for any reason.
5) Do not use it in case of absence of water in the draining basin.
6) After using it, effect the cleaning in the following way: leave the garbage disposal unit in function for 5/10 seconds
running abundantly water. It is possible to add non-corrosive disinfectant.
10) LIQUIDS ASPIRATOR (OPTIONAL ACCESSORY)
The liquids aspirator must be connected to the water system. For the working, open the tap placed in the back side,
position the aspiration cannula in the container to suck; the draining will happen directly in the liquids recovery basin
connected to a can dedicated to the disposal of liquids.